Summary of Contents for mLINE

Page 1: ...t Service Center Repairs Experienced Engineers and Technicians on staff in our State of the art Full Service In House Service Center Facility InstraView Remote Inspection Remotely inspect equipment be...

Page 2: ...mLINE Instruction Manual Bedienungsanleitung Mode d emploi Manual Usuario Instruzioni d impiego Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...

Page 3: ...seal 7 Calibration nut cover 8 Display 9 Handle 10 Tip ejector 11 ID window 12 ID tag 13 Handle collar 14 Cap set of 5 colours Ersatzteile Spitzenabwurf Konusfilterhalter Spitzenkonus Konuszylinder K...

Page 4: ...satzteile Spitzenkonus geh use Spitzenabwurfs balken Spitzenkonus Kolben Kolbenzu sammensetzung Abdeckung Kalibrierschraube Display Handgriff Spitzenabwerfer ID Fenster ID Aufkleber Griffhalter Abdeck...

Page 5: ...anual 1 Bedienungsanleitung 17 Mode d emploi 33 Manual Usuario 49 Instruzioni d impiego 65 81 Specifications 97 mLINE Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan sc...

Page 6: ...UES 7 9 1 Forward pipetting Fig 5 8 9 2 Reverse pipetting Fig 9 8 9 3 Repetitive reverse technique 9 10 RECOMMENDATIONS FOR GOOD PIPETTING 9 11 MAINTENANCE 9 11 1 Daily cleaning of the outer surface o...

Page 7: ...pipetting can cause RSI The manufacturer is not responsible for RSI or any related diseases caused by prolonged pipetting Safe Cone filters Colour Volume 50 pcs bag Cat No Code Channels Range Incremen...

Page 8: ...ck Colour coding Tip ejector collar Tip cone ID window and tag Finger support Handle Calibration nut Safe Cone filter Tip cone Tip cone housing Tip ejector bar Safe Cone filter Artisan Scientific Qual...

Page 9: ...alibration tool also tube and ID window opener 5 Pipette holder 6 Identification tags 7 Safe Cone filters pipettors 10 l 8 Grease 9 Instruction manual 10 Performance certification in accordance with I...

Page 10: ...stall the holder by pressing it against the edge of the shelf Fig 2 4 Place the pipette onto the holder 6 2 Carousel stand A convenient and ergonomical Carousel Stand Fig 3 for six Biohit mLINE pipett...

Page 11: ...lume range of the pipette Chapter 1 8 2 Sealing and ejecting tips To ensure the maximum accuracy and precision Biohit tips are recommended to be used with mLINE pipettes Before fitting a tip ensure th...

Page 12: ...lter ejector without touching the filters by hand Remove the operating button cap and eject the filter by pressing the operating button down Fig 7 Make sure that the filter is disposed into a suitable...

Page 13: ...p Fig 8 0 1 2 9 2 Reverse pipetting Fig 9 1 Fit the tip onto the pipette tip cone 2 Press the operating button all the way to the second stop 3 Place the tip just under the surface of the liquid 2 3 m...

Page 14: ...tte on its side especially with the tip attached Change the tip cone filter regularly recommendation after 50 250 pipetting cycles Never strike the tip cone against the tip tray when mounting the tip...

Page 15: ...the tip ejector collar 1 counterclockwise and remove it 3 Pipettes m3 m10 m20 m100 m200 m1000 Unscrew the tip cone holder 2 counterclockwise and carefully remove it with the tip cone 3 Pipettes m5000...

Page 16: ...ase has spread evenly 5 Check the performance of the pipette Note It is always necessary to check the performance of the pipette after in house service or maintenance 11 3 Sterilization of the pipette...

Page 17: ...fit the tip onto the tip cone 3 Fill the tip with test water and expel to waste five times to reach a humidity equilibrium in the dead air volume 4 Replace the tip Pre wet the tip by filling it once...

Page 18: ...e and the selected test volume Random error imprecision is the scatter of the dispensed volumes around the mean of the dispensed volume ISO 8655 1 Note Biohit specifications are achieved in strictly c...

Page 19: ...d of the calibration nut located at the backside of the handle with the aid of the calibration tool Fig 11 2 Place the hexagonal head of the calibration tool into the hole of the calibration nut 3 Tur...

Page 20: ...r servicing Pipette out of Incorrect operation Follow instructions established specifications Unsuitable tip Use original tips Calibration altered Recalibrate Operating button Liquid has penetrated Cl...

Page 21: ...OUTSIDE THEIR SPECIFICATIONS CONTRARY TO THE INSTRUCTIONS GIVEN IN THIS MANUAL OR WITH OTHER THAN THE MANUFACTURER S ORIGINAL TIPS Each Biohit mLINE pipette is tested before shipping by the manufactur...

Page 22: ...l 1000 l 0 60 0 20 500 l 0 60 0 20 100 l 1 00 0 40 725080 1 500 5000 l 5000 l 0 50 0 20 2500 l 0 60 0 30 500 l 2 00 0 60 725090 1 1 10 ml 10 000 l 0 6 0 2 5000 l 1 2 0 3 1000 l 3 0 0 6 725120 8 0 5 10...