Biohit eLINE Lite Instruction Manual Download Page 15

14

8.  Select the speed in with 

  (1=Min and 5=Max) 

 

9. Press 

or 

 to con

fi

 rm selection and to display the  speed out

10.  Select the speed out with 

 (1=Min and 5=Max)

11. Press 

 to con

fi

 rm selection 

12.  Now the text “TIP” is blinking on the display. 

Attach the tip

13. Con

fi

 rm the programming by double clicking the START-button 

 

Note:

 Dispenser is now 

fi

 xed to the selected tip volume range

14.  Place tip into the liquid and press 

 START-button   to aspirate

15.  Discard the prime excess by 

 

pressing the START-button 

16.  To dispense, position the tip above the 

 

receiving vessel and press START-button 

 

 

and repeat until the cycle is completed

17.  Finally position the tip to discard 

 

excess volume and click twice 

 the 

START-button 

18.  Now the Dispenser is ready for next 

 pipetting 

cycle

Note:

  It  is  also  possible  to  continue  to  aspirate  and  dispense  the  same 

volume without the empty function. To continue, keep the START-button 

pressed during the last dispensing and within one second the direction of 

the arrow will change. Keeping the START button depressed, place the tip 

into the liquid again and then release the START button to aspirate the next 

sample.

OUT

E

SPEED
IN

TIP SIZE

4.00

 

ml

3

10

ml

2 x

SPEED
OUT

TIP SIZE

4.00

 

ml

3

10

ml

2 x

SPEED
IN

TIP SIZE

4.00

 

ml

3

10

ml

2 x

Summary of Contents for eLINE Lite

Page 1: ...Instruction Manual eLINE Dispenser Dispenser Lite Lite...

Page 2: ...HARGING OF THE DISPENSER 8 4 1 Charging with the Biohit eLINE Charging Stand and Charging Carousel 9 4 2 Charging through direct charging system 10 4 3 Charging Recommentations 10 4 4 Electrical speci...

Page 3: ...ior construction of the Biohit eLINE Lite Dispenser enable maximum liquid handling performance with high levels of accuracy and precision The eLINE Lite Dispenser uses positive displacement principle...

Page 4: ...Starter Kit includes 0 5 ml 20 pcs 1 0 ml 20 pcs 2 5 ml 20 pcs 5 0 ml 20 pcs 10 0 ml 20 pcs 1 adapter included Note The tips are disposable do not reuse them 3 QUICK START 3 1 Unpacking The Biohit eL...

Page 5: ...lastic protector Fig 1 1 Push the charging buttons simultaneuously in order to open the battery cover 2 Remove the plastic protector Note It is recommended to charge the pipette for 12 hours before in...

Page 6: ...of the Biohit eLINE Lite Dispenser is performed using the display keyboard operating buttons and tip Fig 2 and 3 Keyboard Display Start button Tip ejection buttons Tip connector Biohit Dispenser Tip...

Page 7: ...ple dispensing mode with Tip Guide The Tip Guide is a unique built in guide that selects the optimal tip size according to selected volume and number of dispensings 3 3 3 Keyboard and operation button...

Page 8: ...cted operating mode Only a soft touch is required to operate the button 3 3 6 Tip Attachment and Ejection Tip attachment 1 Push the tip into the tip connector Make sure that the tip is properly attach...

Page 9: ...e dispenser Please note that if enter is not pressed the dispenser will not start charging even placed in charging stand connected to AC adaptor Note In case you continue to work with the dispenser af...

Page 10: ...adaptor to the electrical outlet check that its voltage setting corresponds to the voltage setting of your electrical outlet Note The Charging Carousel is used to charge the dispenser without the tip...

Page 11: ...battery cover is properly closed 3 Connect the AC adaptor to the electrical outlet as well as to the socket at the back of the dispenser 4 The charging time is 6 hours for the completely discharged b...

Page 12: ...ltages and main plug according to local requirements Output voltage 9 VDC 360 mA AC adaptor for Charging Carousel Input voltages and main plug according to local requirements Output voltage 9 VDC 1200...

Page 13: ...uots Select the tip size Tip Guide Shows the best tip option If there are also other tip options available the TIP SIZE text is blinking View the other options by pressing the Arrow up down key Do you...

Page 14: ...t is attached 2 Select the desired aliquot volume by pressing Note When holding down the the volume display start to scroll faster 3 Press to confirm the selection 4 Select the number of aliquots by p...

Page 15: ...tip above the receiving vessel and press START button and repeat until the cycle is completed 17 Finally position the tip to discard excess volume and click twice the START button 18 Now the Dispenser...

Page 16: ...ith aqueous liquids and liquids with higher viscosity However it is recommended always to choose the tip with the smallest applicable volume in conjunction with high viscosity liquids Please note the...

Page 17: ...ufacturer The use of any other type of battery will immediately void the warranty 7 TROUBLESHOOTING The Biohit eLINE Lite Dispenser have a onboard monitoring program to control the performance of each...

Page 18: ...se original tips Clean the tip connector and attach new tip Dispenser out of the established specifications Unsuitable tip Instrument damaged Use original tips Return to your Biohit service representa...

Page 19: ...ENANCE A OF SERVICE OR NEGLIGENCE OF REGULAR MAINTENANCE AND SERVICE ACCIDENTAL DAMAGE INCORRECT STORAGE OS USE OF THE PRODUCTS FOR OPERATIONS OUTSIDE THEIR SPECIFICATIONS CONTRARY TO THE INSTRUCTIONS...

Page 20: ...10 1 00 1 00 0 50 2 00 1 100 0 2 2 200 0 4 200 20 1 00 1 70 0 50 2 50 1 100 0 5 5 500 1 500 50 0 80 1 00 0 60 3 00 1 100 1 0 10 1000 2 1000 100 0 50 0 50 0 20 0 60 1 100 2 5 25 2500 5 2500 250 0 60 0...

Page 21: ...ectrical and electronic equipments this device must not be recycled as unsorted municipal waste Instead this device must be collected separately in accordance the local recycling regulations The solid...

Page 22: ...www biohit com Due to Biohit s continuing R D effort specifications may change without prior notice Biohit Proline innovations are covered by multiple patents including U S Pat 5 343 769 U S Pat 5 50...

Reviews: