BioFluidix BioSpot BT600 User Manual Download Page 8

Manual - BioSpot BT600 workstation 

 

 

page 7 of 15 

indication for power supply (illuminates while the BioSpot BT600 is switched on and 
connected correctly to the power supply) 

dispense LED red/ yellow 

indication for a dosing event (illuminates at each dispense) 

move LED blue 

indication for axes movement (illuminates while any axis x/y/z is moving) 

power toggle switch (back side) 

enables the power supply of the BioSpot BT600. Using this switch supplies all components 
with power, including PC and dosing electronics. 

emergency switch  

Pressing the emergency switch disconnects all internal components from the power supply, 
except the internal PC and the power switch on the back side. The emergency switch can be 
used to stop non-intended movement of one of the axis or non-intended dosage. 

PLEASE NOTE:  

After using the emergency switch, the axes need to be re-referenced! 

Further, it is required to re-boot the internal PC after pressing the emergency switch in order 
to put the BioSpot back into operation. 

 

 

Dispenser connection 

Connector sockets for up to 8 PipeJet dispenser units; U = 0 to 180V 

 

Warning: 

High voltage hazard at plugs and sockets!

 

USB connector 

Standard USB 2.0 connector, intended for the connection of a TopView camera  
(may be ordered separately) 

Connection for ring-light 

Current controlled connector for the TopView camera illumination 
(may be ordered separately) 

Summary of Contents for BioSpot BT600

Page 1: ...BT BioFluidix GmbH Engesserstra e 4a D 79108 Freiburg Germany www biofluidix com User Manual BioSpot BT600 Version 1 1 Date March 2018...

Page 2: ...ist of parts 5 5 Description of the BioSpot BT600 6 5 1 BioSpot BT600 workstation 6 5 2 Connectors at the control electronics located at the z axisFehler Textmarke nicht definiert 5 3 Backside connect...

Page 3: ...es on cables or damages on mechanical parts If it is supposed that it is not possible to launch a safe operation you have to take it out of service and protect it against an unintended operation It is...

Page 4: ...into the BioSpot BT600 device It is prohibited to store or place any object on the platen of the BioSpot BT600 which is not in the purpose of use Keep your hands away from all moving elements during o...

Page 5: ...a laboratory or industrial environment only the BioSpot BT600 is to be operated by skilled personnel who are familiar with the user s manual only The BioSpot BT600 workstation without installed compo...

Page 6: ...e in outdoor environment Use the device indoors in clean and dry environment The installation base needs to offer a stationary even and robust surface It must resist the appearing acceleration forces...

Page 7: ...igure 1 Photograph of the BioSpot BT600 workstation indicating the most important functional parts Dispenser connection Plug socket to connect a dispenser e g up to eight PipeJet dispensers Warning Hi...

Page 8: ...mponents from the power supply except the internal PC and the power switch on the back side The emergency switch can be used to stop non intended movement of one of the axis or non intended dosage PLE...

Page 9: ...re essential methods to improve or support the development of the desired dosing process or be applied to implement process control Possible accessories are listed and explained in the following TopVi...

Page 10: ...control by correlation of generated signals Pressure controlled reservoir o guarantees constant pressure at the dosing nozzle o improves long term stability of the dosing process o enables big reservo...

Page 11: ...socket 90 240V o Fuse 6A B C characteristic 3 Make sure the emergency switch is unlocked 4 Switch the backside power toggle switch to 1 on to start the BioSpot BT600 o the internal PC will start up au...

Page 12: ...into the substrate holder Move reservoir holder down and fix it with the allen key Connect Tygon ND 100 65 to the intermediate reservoir using the black rubber adaptors and the Serto aspiration connec...

Page 13: ...on of the device you can request for the original set of registry entry by BioFluidix 7 2 Interfaces Ethernet socket Typ RJ45 Transfer rate 10 100 1000Mbit s USB 2 0 USB connectors DVI connector Conne...

Page 14: ...nsions w x l x h 550 mm x 636 6 mm x 539 2 mm Weight 32 kg depending on the individual configuration Environmental operation conditions Temperature 10 C 30 C Humidity 10 80 non condensing conditions 7...

Page 15: ...ispensing the BioSpot BT600 workstation check for contaminations of hazardous or environmental and health threatened substances that may have been formed during use and remove them if required Dispose...

Page 16: ...stions or comments please contact BioFluidix GmbH Engessertsra e 4a D 79108 Freiburg Germany phone 49 761 458938 0 fax 49 761 458938 99 e mail info biofluidix com Internet www biofluidix com Freiburg...