GeneExplorer Thermal Cycler Operation Manual
Chapter Four Operation Instructions
-
15
-
Interface 1-1
Interface 1-2
2.4 Extended parameter
On the interface of PCR program setting, click “
Extended parameter
” and it will prompt a box
for setting extended temperature, extended time, speed, and gradient and other operations.
Extended temperature: the volume of temperature increase or decrease for each loop. “-/+”
changes the positive and negative of data.
Extended time: time increase for each loop.
Speed: ramp rate
Gradient
: gradients temperature refers to the temperature difference between the most
left and most right ranks of the module. In “Temperature”, the temperature of the most
left rank is set. GE-48DG, GE-96G, GE-384G own this function.
Ladder Temp:
It refers to the temperature difference between two adjacent modules.
The range is 0.1℃~6℃. “Temperature” is set to the temperature of the leftmost module.
GE-48DS, GE-96S own this function.