3. Procedure to adjust an isotope dial setting in the 

Atomlab 500 dose calibrator 

1.

Press the HOME key

2.

Press SELECT UTILITIES

3.

Select DOSE CALIBRATOR CONFIGURATION

4.

Select ISOTOPE LIST – you must now enter the password

5.

The ISOTOPE LIST screen is now displayed

6.

Select CHANGE DIAL VALUES; the Change Dial Values Isotope List is now displayed

7.

Use 

⇑⇓

arrows to go to the NEW ISOTOPES LIST

8.

Select ISOTOPE Ra-223

9.

Select CHANGE DIAL VALUE

10. Enter dial value you determined in Part II
11. Select ENTER
12. Select ENTER to save the new dial value or CANCEL to cancel the change
13. Select HOME to return to the Home Screen

NOTE:

You have now adjusted Ra-223 dial value on your isotope list.

4. Procedure to change one of the button isotopes to be Ra-223 

1.

Press the HOME key

2.

Press SELECT UTILITIES

3.

Select DOSE CALIBRATOR CONFIGURATION

4.

Select ISOTOPE LIST – you must now enter the password

5.

The ISOTOPE LIST screen is now displayed

6.

Select the SETUP Button isotopes  

7.

Use the 

⇑⇓

arrows to scroll down thru the routine isotopes and continue until you 

display the new isotopes  

8.

Highlight the Ra-223 isotope by pressing it

9.

Select the button you wish to change to be the Ra-223 – you will see that the name on 
the button changes to the Ra-223  

10. Select ENTER to record this as a change and return to the isotope list choice screen – 

the system is now set for Ra-223 to be used as a button isotope

11. Select HOME to return to the Home Screen