ATP Calibrator Set 

 
 

 

GENERAL DESCRIPTION

 

The calibration devices in this kit are used for performance 
verification and calibration of the LIGHTNING MVP ICON

 

and LIGHTNING MVP

®

 systems. The luminescent calibrator 

consists of a 

14

C radioactive source that emits a very low level 

(5 x 10

5

 d.p.m., or 0.225 microcuries) of low energy ß radiation 

in a plastic scintillation matrix. This matrix is configured to give 
constant light output, within 10% of its original value, at a 
sufficiently high level to make consistent verification and 
calibration functions possible over the five-year life of the 
calibrator. The non-luminescent black plastic calibrator provides 
a reliable low-end or “dark” source reading. 

Part No. 64012-00 

KIT COMPONENTS 

One luminescent calibrator 

One non-luminescent calibrator 

RECOMMENDED USE

 

A.

 

Performance Verification:

 If a given test point result 

or LIGHTNING MVP ICON ATP Positive Control result 
is not what is expected, it is suggested that the calibrators 
be used. You may wish to perform daily or weekly 
performance verification checks consisting of simply 
inserting and reading the calibrators. Calibrator results 
should be compared with the Zone value stated on the 
calibrator label and with previous calibration results.                
If the calibrator results are within specifications (see                          

INTERPRETING RESULTS

), you can be confident 

that the MVP ICON or MVP is functioning properly. 

B.

 

Calibration: 

The instrument’s Zone readings are 

standardized by a two-point calibration procedure.  

Note:

 All MVP ICON and MVP instruments are factory 

calibrated prior to being shipped. Because several factors can 
cause the calibrator results to be out of specification, a full 
calibration should be performed only if either of these two 
conditions are true, and all of the conditions in “Causes of 
Inconsistent Calibrator Readings” have been ruled out: 

 

(a)

 

Luminescent calibrator readings are consistent within 0.1 
Zones of each other but are not within 0.1 Zones of the 
Zone value printed on the calibrator label. 

(b)

 

Non-luminescent calibrator readings are consistent within 
± 0.3 Zones of each other but are not within the 1.0 to 
1.5 range. If, after cleaning and repeated tries, consistent 
calibrator results cannot be obtained, call BioControl 
Technical Support at 800.245.0113. 

Note: 

  Accepting a particular calibration will set the sensitivity 

for all subsequent readings. It is not possible to change a 
calibrated instrument back to a former calibration. 

 

 

STORAGE 

Store at 15 – 25 

˚

C in a tightly closed container, away from 

any light sources. 

MVP ICON TEST PROCEDURES 

The calibrators are inserted and read in the MVP ICON and 
MVP in the same manner as sampling devices. Unlike the 
sampling devices, the calibrators do not require activation.  
DO NOT ATTEMPT TO ACTIVATE THE LUMINESCENT 
CALIBRATOR. 

Hold the calibrator by the wider end with two flat sides and 
insert the opposite end into the MVP ICON or MVP (an arrow 
on the luminescent calibrator’s label indicates which end is to 
be inserted into the instrument). When inserting into the MVP 
ICON, ensure the flat sides of the calibrator’s base chamber 
are facing left and right. 

A.

 

 MVP ICON Performance Verification 

(a)

 

Select “Device” from the bottom menu and ensure the 
MVP ICON is in ATP mode. 

(b)

 

Select “Luminescent Cal” or “Non-Luminescent Cal”,             
as appropriate, to run the verification. 

(c)

 

Follow prompts. 

B.

 

Calibrating the MVP ICON 

(a)

 

Select “Device” from the bottom menu and ensure MVP 
ICON is in ATP mode. 

(b)

 

Select “Calibrate” and enter the reference value located 
on the luminescent calibrator label.

 

 

(c)

 

Insert the luminescent calibrator and press “Calibrate.” 
Repeat with the non-luminescent calibrator. 

(d)

 

The MVP ICON will display “Calibration Successful” when 
properly calibrated and will store the calibration in the 
device. The last calibration date will appear on the 
homescreen of the device. If the “Calibration Failed” 
message appears, the calibration will need to be repeated. 
If failed message still appears, contact BioControl 
Technical Support at 800.245.0113. 

LIGHTNING MVP TEST PROCEDURES 

A.

 

LIGHTNING MVP Performance Verification   

(a)

 

Open the sample chamber and insert the calibration 
device. Close the sample chamber. 

(b)

 

Select “Read ATP” from the main menu. 

(c)

 

It is suggested that luminescent and non-luminescent 
calibrator results be stored in the MVP memory for later 
retrieval. Assigning these results test point numbers will 
facilitate data collection and retrieval: for example, 996 
non-luminescent, 997 luminescent. 

(d)

 

Assign a test point number (if desired) and press the 
“Enter” key to begin reading. 
 

 
 
 