5.4   Attach the Reservoir:

There is a 60 cc syringe in the accessory kit.  Pull the plunger out and connect it to the luer-loc 

connector at the end of the blue tubing on the right side of the box (pull the tubing out first).  Slide the 
flanges on the top of the syringe barrel under the screw heads on either side of the notch in the right 
wing.  Fill the reservoir with 

FILTERED 

water or buffer, depending on your needs.  Small particles of 

dust and debris can clog the valve in the piston tip.  The salt in buffer will reduce the carryover between 
fractions and keep your samples isotonic.  However, it will also put more load on your scintillation 
cocktail if you are counting your rinses.

11

Summary of Contents for GRADIENT STATION 153

Page 1: ...ftware version 8 03 Gradient forming software version 5 3 OPERATOR S MANUAL 2013 BioComp Instruments Inc 650 Churchill Row Fredericton New Brunswick Canada E3B 1P6 Ph 800 561 4221 506 454 6410 Fx 506...

Page 2: ...ient Station 7 3 2 Gradient Forming attachment 8 SECTION 4 ORDERING INFORMATION 4 1 The Gradient Station Base unit 8 4 2 Gradient Forming 8 4 2 1 Tube holders 8 4 3 Gradient Fractionation 8 4 3 1 Tube...

Page 3: ...PART 2 GRADIENT MASTER MANUAL PART 3 GRADIENT FRACTIONATOR MANUAL PART 4 GRADIENT PROFILER MANUAL where applicable PART 5 GRADIENT RUN PARAMETERS last pages...

Page 4: ...this period cannot be honored 1 2 MAINTENANCE Other than the tightening of loose screws and the removal of spilled gradient solutions no user service is required The instrument contains a computer whi...

Page 5: ...r Coombs work with the purification of viruses led him to develop an exquisitely sensitive band visualization system T4 phage at 108 phage ml can be visualized with the naked eye Ribosomal subunits ar...

Page 6: ...l tube length Number of Fractions 1 99 in each of 10 sections per gradient Band Illumination System 20W halogen spot with dichroic reflector Rinse Capability programmable in any order up to 9 events b...

Page 7: ...SECTION 3 PARTS DIAGRAMS 3 1 THE GRADIENT STATION 7...

Page 8: ...ailable for quick seal vertical rotor tubes but they cannot be fractionated with this device 4 3 Gradient Fractionation Includes the following accessories 1 Manual 1 Left wing for extended working are...

Page 9: ...with dichroic reflector 151 905 check valve assembly for piston air rinse sample 151 905 11 check valve assembly for piston air rinse sample for 11 mm tip 151 931 Bag of 50 mist filter discs 151 919...

Page 10: ...racks 5 3 Assemble the Gradient Forming attachment Carefully remove the Gradient Forming attachment from its box and attach it to the right wing position While supporting it screw in one of the same...

Page 11: ...the top of the syringe barrel under the screw heads on either side of the notch in the right wing Fill the reservoir with FILTERED water or buffer depending on your needs Small particles of dust and...

Page 12: ...d and two 6 pin sockets labelled tilt and rotate Insert the motor plugs onto their 6 pin sockets Note that the motor plugs have red stripes on one side The must face OUT away from the white plastic ba...

Page 13: ...he computer to the bent edge of the front panel using 3 more 8 32 screws as shown 5 6 Install the power plug You are now ready to power up the unit Find the external power supply 12VDC Powergate trans...

Page 14: ...ns that its smart key has been pressed The dedicated keys in the second row are used only for fractionation and have self explanatory labels AIR turns on the air pump forcing air into the collection t...

Page 15: ...hown in A Adjust the feet under the corners of the instrument until the level is centered Next turn the level so it runs side to side as shown B Turn the knob clockwise or counter clockwise to level t...

Reviews: