BioCision XT Starter Instructions For Use Download Page 2

When using  XT Starter to cool samples in a CoolRack using a charged XT Cooling Core

•  Remove XT Cooling Core from the freezer and place into holder.

•  Place CoolRack or CoolSink module directly onto core and allow to equilibrate to 4

o

C   

(approximately 10-15 minutes). 

•  Load samples.

When using a pre-chilled (4

o

C) CoolRack or CoolSink module

•  Remove XT Cooling Core from the freezer and place on the benchtop for approximately 10 minutes. When the 

temperature indicator displays 1

o

C, XT Cooling Core is ready to use. 

IMPORTANT: Failure to allow XT Cooling Core to reach 1

o

C may result in sample freezing.

•  Place XT Cooling Core into holder.
•  Place the pre-chilled CoolRack or CoolSink module of choice onto the core.
•  Load samples.

Care and Cleaning

The XT Starter holder is constructed from a cross-linked closed-cell dense polyethylene foam. The material has 
excellent resistance to fluid absorption and abrasion. Maximum temperature exposure: 60°C.  Avoid prolonged 
exposure to UV light sources. All components (XT Holder and XT Cooling Core) can be cleaned with aqueous 
detergents, alcohol, 10% bleach, and acid/base viricide (such as Virkon S) solutions. Rinse with clear water after 
using cleaning solutions. Do not autoclave.

When using  XT Starter to cool samples in a CoolRack using a charged XT Cooling Core

•  Remove XT Cooling Core from the freezer and place into holder.

•  Place CoolRack or CoolSink module directly onto core and allow to equilibrate to 4

o

C   

(approximately 10-15 minutes). 

•  Load samples.

When using a pre-chilled (4

o

C) CoolRack or CoolSink module

•  Remove XT Cooling Core from the freezer and place on the benchtop for approximately 10 minutes. When the 

temperature indicator displays 1

o

C, XT Cooling Core is ready to use. 

IMPORTANT: Failure to allow XT Cooling Core to reach 1

o

C may result in sample freezing.

•  Place XT Cooling Core into holder.
•  Place the pre-chilled CoolRack or CoolSink module of choice onto the core.
•  Load samples.

Care and Cleaning

The XT Starter holder is constructed from a cross-linked closed-cell dense polyethylene foam. The material has 
excellent resistance to fluid absorption and abrasion. Maximum temperature exposure: 60°C.  Avoid prolonged 
exposure to UV light sources. All components (XT Holder and XT Cooling Core) can be cleaned with aqueous 
detergents, alcohol, 10% bleach, and acid/base viricide (such as Virkon S) solutions. Rinse with clear water after 
using cleaning solutions. Do not autoclave.

12 E. Sir Francis Drake Blvd, Suite B, Larkspur CA, 94939 USA  [email protected]   www.biocision.com

XT Starter Instructions, January 2014 

©2014. BioCision, LLC. All rights reserved. Patents pending. BioCision, Standardizing Samples designations are trademarks owned by BioCision, LLC. 

15010

 Rev A 

12 E. Sir Francis Drake Blvd, Suite B, Larkspur CA, 94939 USA  [email protected]   www.biocision.com

XT Starter Instructions, January 2014 

©2014. BioCision, LLC. All rights reserved. Patents pending. BioCision, Standardizing Samples designations are trademarks owned by BioCision, LLC. 

15010

 Rev A 

Item No.

Description

Dimensions (L x W x H):

BCS-504

XT Starter, purple

Exterior: 16.2 x 12.0 x 5.5 cm / 6.4 x 4.8 x 2.2 in

BCS-513

XT Holder, purple

BCS-511

XT Cooling Core 

Item No.

Description

Dimensions (L x W x H):

BCS-504

XT Starter, purple

Exterior: 16.2 x 12.0 x 5.5 cm / 6.4 x 4.8 x 2.2 in

BCS-513

XT Holder, purple

BCS-511

XT Cooling Core 