4. Read the result.

The measurement result will be done in 

3 seconds.

The reading is shown together with green

LED and three short beeps.

 The body of the thermometer is not water-resistant.

     Never put the thermometer under a running tap or submerge it into 

     water. Use a soft and dry cloth to clean it. Do not use abrasive  

     leaners.

 Store the thermometer in a cool and dry location. Free from dust and

      away from direct sunlight.

Remove the battery

The thermometer comes with two 1.5 V AAA alkaline batteries. The 

meter will display  “           ” to alert you when the meter power is getting 

low, please follow the steps below to replace new batteries

immediately.

1. Remove the battery cover as the arrow 

    direction accordingly.

2. Remove the old batteries and replace 

     with two 1.5V AAA size alkaline batteries.

      

          Positive ( + ) and Negative ( - ) are displayed 

        

on the back of battery cover.

      

3. Close the battery cover.

NOTE :

1. Although the thermometer works when "            " appears,we still   

    recommend that you change the batteries to obtain an accurate 

    result.

2. Remove the batteries if stored for a long period of time.

3. The batteries should be kept away out of child’s reach. If they are 

    swallowed, promptly see a doctor for help.

METER MEMORY

Recalling the Memory

Your thermometer stores 20 most recent 

readings

1. Be sure the thermometer is OFF before 

   recalling this memory.

2. Press Memory Button to enter the 

    memory mode.

Each time you press the Memory Button, a 

result will be displayed

in the order of dates (latest result

shown first), together with ”         ”

and number ( from 1 to 20 ).

LED in green or red will appear according

to the memory reading.

When the memory is full, the oldest result is

deleted as the new one added. When the 

last record displayed in the Display, press

Memory Button again to return the first 

record.

3. Exit the memory.

    

Press Set Button to exit the memory or 

    keep the meter in idle for 10 seconds to 

    switch off automatically.

MAINTENANCE

Care & Cleaning 

● 

 

The probe is not waterproof. Please wipe with a clean and dry 

      cotton swab to clean the probe on the inside.

SYSTEM TROUBLESHOOTING

Measurement Result

Troubleshooting

The table below shows problems you may encounter. All error 

messages below would be shown together with backlight. Please 

follow "what to do" to resolve problems.  If the problem still exists, 

please call your local dealer for help.

SPECIFICATIONS

Model No.: 1265

Dimension & Weight : 151.5 x 55 x 41 mm, 74g (includebatteries)

Power Source : 2 x 1.5 V AAA alkaline batteries

Displayed temperature range :

Forehead : 32°C to 43°C (89.6°F to 109.4°F)

Object : 0°C to 100°C (32°F to 212°F)

Display resolution: 0.1 °F/0.1°C

Accuracy :

FOR BODY MODE:

±0.2°C (±0.36°F) from 36.0°C (96.8°F) to 39.0°C (102.2°F)

±0.3°C (±0.54°F) from 32.0°C (89.6°F) to 35.9°C

(96.6°F) and from 39.1°C (102.4°F) to 43.0°C ( 109.4°F)

FOR OBJECT MODE : ± 1°C (±2°F) from 0°C (32°F) to 100°C (212°F)

Temperature unit : °C or °F

Operating temperature range : 16°C to 40°C (60.8°F to 104°F)

Operating humidity : 95% RH or less

Storage temperature range : -25°C to 55°C (-13°F to 131°F)

Storage humidity : 95% RH or less

Memory capacity : 20 measurements

Expected service life : 3 years

DISPOSAL

Adhere to the applicable regulations when disposing of the device.This product must not be disposed of

 together with domestic waste.All users are obliged to hand in all electrical or electronic devices, regardless 

of whether or not they contain toxic substances, at a municipal or commercial collection point so that they

 can be disposed of in an environmentally acceptable manner.

Please remove the batteries before disposing of the device/unit. Do not dispose of old batteries with your 

household waste, but at a battery collection station at  a recycling site or in a shop.

REFERENCE STANDARDS

Device Standard :

Device Corresponds to the requirements of the

standard for infrared thermometers

ASTM E1965-98

EN 12470-5

EN 60601-1-2 : 2007/AC:2010

Electromagnetic Compatibility : Device fulfills the stipulations of the standard EN 60601-1-2

The stipulations of EU-Directive 93/42/EEC for Medical Devices Class Ila have been fulfilled.

SYMBOL INFORMATION

DETAILED INFORMATION

About Normal Body Temperature & Fever

The temperature in the forehead and temple area differs from the 

internal temperature, which is taken orally or rectally. Vasoconstriction, 

an effect which constricts the blood vessels and cools the skin, can

occur during the early stages of a fever. In this case, the temperature 

measured by the Biocare Forehead Thermometer may be unusually low. 

If the measurement therefore does not match the patient’s own 

perception or is unusually low, repeat the measurement every 15 

minutes. As a reference, you can also measure the internal body 

temperature using a conventional oral or rectal thermometer.

Body temperature can vary from one individual/person to next. It also 

varies by location on the body and time of day. Below shows the 

statistical normal ranges from different sites. Please keep in mind that 

temperatures measured from different sites, even at the same time, 

should not be directly compared. Fever indicates that the body 

temperature is higher than normal. This symptom may be caused by 

infection, overdressing or immunization. Some people may not 

experience fever even when they are ill. These include, but are not

limited to, infants younger than 3 months old, persons with compro-

mised immune systems, persons taking antibiotics, steroids or 

antipyretics (aspirin, ibuprofen, acetaminophen), or persons with 

certain chronic illnesses. Please consult your physician when you feel ill

even if you do not have fever.

Table*

1

 Body Site Normal Temperature Range

*

1

. Body Temperature at WebMD; website: http://firstaid.web 

d.com/body-temperature retrieved at 2010 Jan 7.

0.6°C (1°F) or more above or below

37°C (98.6°F)

0.3°C to 0.6°C (0.5°F to 1°F) higher

than oral temperature

0.3°C to 0.6°C (0.5°F to 1°F) lower

than oral temperature

Oral

Rectal/ear

Axillary(armpit)

   

Measured temperature falls outside  

the displayed temperature range of 

32°C to 43°C (89.6°F to 109.4°F), is 

shown on the display with red LED. 

 Hi-temperature ≥43°C (109.4°F)

 Lo-temperature < 32°C (89.6°F)

Measured temperature falls in

normal reading, within the range

of 32°C to 37.5°C (89.6°F to 99.5°F), 

is shown on the display with green 

LED.

Measured temperature falls in

fever reading, ≥37.5°C (99.5°F) and

< 43°C (109.4°F), is shown on the 

display with red LED.

Appear when

environmental

temperature is

below system

operation range.

Put the thermometer

under operating

temperature range of 

16°C to 40°C(60.8°F 

to 104°F)

MESSAGE

WHAT IT MEANS

MESSAGE

WHAT IT MEANS

WHAT TO DO

Appear when

environmental

temperature is

above system

operation range.

Battery is low

and "            " appears

on display. 

Put the thermometer

under operating

temperature range of 

16°C to 40°C(60.8°F 

to 104°F)

Please replace

batteries as

soon as possible.

If the thermometer

is broken or

affected by EMF.

Please contact

the customer

service line.

MESSAGE

SYMBOL

REFERENT

WHAT IT MEANS

WHAT TO DO

Read instructions before use

Manufacturer

Serial number

CE mark

Authorized representative in the European Community 

Caution, consult accompanying documents

Type BF Equipment

Collection for electrical and electronic equipment

Shanghai International Trading Corp. 
GmbH(Hamburg)
Eiffestrasse 80, 20537 Hamburg, Germany
TEL:+49‐40‐2513175
FAX:+49‐40‐255726

Biocare Asia Co., LTD
No.260,Mayun Road, New District Suzhou,
Jiangsu,215129,P.R.C
TEL:+86‐512‐68091772  
FAX:+86‐512‐68097761 
www.bca-medical.com

EN 60601-1-4 :1996

EN 60601-1-6 : 2010

Reviews: