Battery charging time

4 hours when powered off

Use- life

3 years

Shelf- life

2 years

Charger Isolation: class ll 

double isolation

5V AC/DC Adapter

AC Power for battery 

charger

100-240V, 50-60 Hz, 10VA 

max

Battery for Patch

Capacity

5 days of continuous use

Shelf- life

3 years

Non-rechargeable

One-time (single) use

Environmental
Operating temperature

4°C to 39°C (39°F to 103°F)

Operating humidity 

Up to 95%, non-condensing

Pressure

700 to 1060 hPa

Operating altitude

-378 m to 3050 m
(-1240 feet to 10000 feet)

Storage and transportation
Storage temperature

-20°C to 70°C (-4°F to 158° F)

Humidity

Up to 95%, non-condensing

Pressure

465 hpa to 1060 hPa

Operating altitude

-378 m to 6098 m 

 

(-1240 feet to 20000 feet)

Physical Characteristics - Wristwatch

Dimensions (monitor 

enclosure)

48 x 38 x 16 mm

Weight

2 oz. (55 g) including battery

Physical Characteristics - Patch

Dimensions

40 x 40 x 16 mm

Weight

Weight 1 oz. (30g)

Compliance

Equipment Classification

IEC 60601-1

Summary of Contents for BB-613WP

Page 1: ...Monitoring Wristwatch Disposable Patch User Guide...

Page 2: ...or transmitted in any form by any means electronic mechanical photo reproductive recording or otherwise without the express prior written permission of Biobeat Technologies Ltd Biobeat Technologies L...

Page 3: ...or monitoring of vital signs in clinical and non clinical settings Features and benefits Easy of use Ability to transmit the data to the App BLE Communication 3 CONDITIONS FOR USE 3 1 Indications A ba...

Page 4: ...N 60601 1 2 ED 4 for EMC of medical equipment The device has Class BF III compliance 4 3 Safety Instructions Warnings DO NOT USE BEFORE READING THIS USER MANUAL Only apply the device on clean intact w...

Page 5: ...unit ESU This device should not be used adjacent to other equipment If adjacent use is necessary the device should be observed carefully to verify normal operation The device is intended for indoor o...

Page 6: ...Do not use an adapter that was not supplied with the device The device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin Factors that may degrade pulse ox...

Page 7: ...rformance of this device Medical electrical equipment needs special precautions regarding EMC and all equipment must be installed and put into service according to the EMC information specified in thi...

Page 8: ...eat device 5 2 Cleaning the Biobeat device Caution Should the device become wet wipe off all moisture and allow sufficient time for drying before operating Caution Contact with solvents can cause seve...

Page 9: ...nicipal waste Contact your local distributor for unit disposal Caution consult accompanying documents SN Serial number Manufacturer Manufacturing date Consult instructions for use Catalogue number Cau...

Page 10: ...R pulse rate 40 240 bpm Blood pressure SYS 60 to 250 mmHg 5 DIAS 40 to 150 mmHg 5 Accuracy Arterial Oxygen Saturation 70 to 80 80 to 90 90 to 100 3 Pulse Rate Accuracy 3 bpm Resolution Arterial Oxygen...

Page 11: ...F Operating humidity Up to 95 non condensing Pressure 700 to 1060 hPa Operating altitude 378 m to 3050 m 1240 feet to 10000 feet Storage and transportation Storage temperature 20 C to 70 C 4 F to 158...

Page 12: ...Devices transmit information tracked by the applicable Biobeat Device Biobeat s Application allows the User the User to use Biobeat Devices functionalities as specified under the Documentation as defi...

Page 13: ...itled to change these Terms from time to time including in order to comply with applicable law and or regulation by posting an updated version on Biobeat s Application Any such change shall be effecti...

Page 14: ...Website without limitation use of any device that is not manufactured distributed or sold by or on behalf of Biobeat including without limitation any knock off or counterfeit versions is prohibited ii...

Page 15: ...ower supply included in with Biobeat Devices It is further clarified that Biobeat does not provide any warranty regarding the installation maintenance or service of Biobeat Devices and that the Warran...

Page 16: ...ckups services or treatments by physicians and other medical care professionals Additionally Biobeat makes no warranty that Biobeat Devices and or the Biobeat Services will meet the User s requirement...

Page 17: ...ABOVE BIOBEAT SHALL HAVE NO LIABILITY WHATSOEVER FOR ANY INACCURACY PERFORMANCE FAILURES UNAVAILABILITY AND OR LOSSES ARISING OUT OF ANY USE OF BIOBEAT DEVICES AND OR BIOBEAT S SERVICES NOT IN ACCORD...

Page 18: ...the User and Biobeat Biobeat shall be the sole and exclusive owner of all rights to any and all information which was absorbed collected and accumulated by Biobeat Devices or provided to Biobeat and...

Page 19: ...will be governed by and construed in accordance with the laws of the State of Israel The User hereby irrevocably agree that the competent courts of the Tel Aviv Jaffa District of Israel shall have ex...

Page 20: ...Biobeat Technologies Ltd Email Info bio beat com Website http www bio beat com UM 0001 03 March 2020...

Reviews: