22
password modification interface. Enter the original password to enter the modification interface to
modify the user password and click “confirm” button to exit.
Glass door up button: press the up button continuously, the glass door will continue to rise, and stop
when it is 200 mm away from the worktable, continue to press the up button until the lowest point of
travel is reached, and release the button, the glass door will stop moving.
4.4. Instructions for Operation
a
、
Connect the same power reply, as required of equipment
b
、
Open the power lock, LCD display lights up and alarm rings at the same time, then the machine
enters to standby status. Wait for the operator to input button to use it.
c
、
Press POWER button, then the following functions are available: Fluorescent lamp, UV lamp,Fan,
Mute, Sockets, Front window up and down, Reservation timing
When front window is closed and fluorescent light is off, then press the UV button to
select the sterilization function.
When UV lamp working avoid looking straight.
d
、
Please move the front window at 200mm height from the worktable, turn on the fan, make sure the
experiment should be started after fan working for at least 10 minutes.
(1) For operating safety, please put testing materials inside the cabinet in advance, and
keep the front window at 200mm height from the worktable during operation.
(2)UV lamp intensity should be tested regularly. If there is no test conditions, it should be
replace when the UV timer on the display indicate the working time reaches to 1000 hours.
e
、
Please move the front window at 200mm height from the worktable, turn on the fan, make sure the
experiment should be started after fan working for at least half an hour.
For operating safety, please put testing materials inside the cabinet in advance, and keep
the front window at 200mm height from the worktable during operation.
After finishing the experiment, please move the front window down to the bottom, and make sure to
sterilize the cabinet by UV lamp for 30 minutes before turning off the cabinet.
4.5 Daily maintenance
Because the operating time will directly affect the judgment of maintenance needs, we recommend
the user keep a detailed record of operating time for reference.
When doing maintenance, please pay attention to cut off the power, so as to avoid electric
shock!
4.5.1 Preparations before maintenance
Soap, hot water or warm water, a soft cotton cloth, dry cloth or towel, medical alcohol or other
disinfectants, 100 dilution of household bleach, abrasive household cleaners, sterile water
Before fumigation should not open the cabinet enclosure, prohibit removing anyparts
from the Microbiological Safety Cabinet!