2.2 Module structure and function
2.2.1 Structure of magnetic rod and rod sleeve
Figure 3-1
The magnetic rod module and the rod sleeve module are independent and independently controlled
by the dual Z-axis motors. The rod sleeve module can run independently during the shake and lysis
process, and the magnetic rod module and the rod sleeve module also can run as a whole during
adsorption process. The motors do not interfere with each other during running.
2.2.2 Function of plate positions
Figure 3-2
There are two positions in the instrument where you can put the deep hole plate, a total of 24
columns of hole positions, of which:
Columns 1-8 identified in Figure 3-2 (the actual hole positions are columns 1, 6, 7, 12, 13, 18, 19, 24)
are heating modules, and the temperature control range is from room temperature to 100°C.
Four leaf springs are placed on both plate positions to fasten the deep hole plate.
2.2.3 Function of floodlight
This instrument is equipped with LED floodlights to observe the operating conditions in the cabin
during running.
Light on/off method: After the instrument is turned on, there is a "Light" button in the upper right
corner of the software interface (the position of the button will not be changed in any interface).
Click the "Light" button to turning on or turning off the light in any state.