BIO RAD PowerPac User Manual Download Page 14

Problem

Cause Solution

E8 error code

Regulation error

Turn power supply off, then on to

reset.
displayed

If problem persists, contact Bio-Rad
Technical Resources

E9 error code

Change in Load 

Check and correct any potential 

displayed

Resistance. 

resistance problem then,
• press run/pause key to resume run

or,

The PowerPac 

• press any key, other than the

run/pause

Basic detects drastic

key, to clear the error code.

changes in resistance 
which may indicate
failure of the cell’s power
leads or a loose output
connection.

Note: Certain applications 
exhibit fluctuations in 
resistance that can 
trigger the change in 
resistance error code.
If this is the case, the 
change in resistance 
feature can be deactivated 
to allow uninterrupted 
completion of a run. 
(see Section 3, Step 8)

Caution:  Deactivating this 
safety feature increases the 
possibility that a failure of the
power leads or loose output
connection will not be 
detected.

E10 error code

Unacceptable value(s) 

Clear the code by pressing any key 

displayed

entered 

other than the run/pause key.
Then, enter values within range of
PowerPac Basic and press the
run/pause key

E12 error code

Internal Over Current 

Possible power supply malfunction.

displayed
E13 error code

Internal Short Circuit

Check for and correct problems such

displayed

as dirty contacts, frayed wires, 

E14 error code

Internal Over Voltage

excessive buffer concentration. Then 

displayed

press any key other than the run/pause 

E15 error code

Internal Short Circuit

key to clear the code.

displayed
E16 error code

Hardware failure

Contact Bio-Rad Technical Resources

displayed
E17 error code

Hardware failure

displayed
E98 error code

Internal system error

displayed
E99 error code

Internal system error

displayed

8

Summary of Contents for PowerPac

Page 1: ...PowerPac Basic Power Supply Instruction Manual Catalog Number 164 5050 For Technical Service Call Your Local Bio Rad Office or in the U S Call 1 800 424 6723...

Page 2: ......

Page 3: ...2 Troubleshooting 7 4 3 Replacing a Fuse 9 4 4 Firmware Version Number 9 Appendix A Specifications 10 Appendix BWarranty and Ordering Information 11 List of Figures 1 Front View 2 Rear View 3 Front V...

Page 4: ......

Page 5: ...cold room the unit can be operated immediately However when removing the power supply from the cold room let the unit equilibrate to room temperature for a minimum of 2 hours before using it 6 Never...

Page 6: ......

Page 7: ...s first reached When the set limit of the non constant parameter is reached and the power capability of the unit is not exceeded the power supply will switch making the non constant parameter the new...

Page 8: ...ve the power supply carefully inspect the container for any damage which may have occurred in shipping Severe damage to the container may indicate damage to the power supply itself If you suspect dama...

Page 9: ...s Changes the value of the selected parameter If the Scroll Key is pressed constantly for more than 5 units in either direction the values will increase decrease in increments of 10 to reach the desir...

Page 10: ...arameter Press the Constant key to select the constant parameter either voltage or or current The LED on the parameter key corresponding to the selected constant parameter will light up The display sh...

Page 11: ...for 2 seconds Caution Always use the stop key until the display shows Pfd This indicates to terminate a run in progress Use the Power Failure Detection is activated of the power switch to terminate a...

Page 12: ...tant parameter Note To change from an un timed to a timed run stop the run and re program 12 End of run When a run is completed i e a timed run has ended or an untimed run is stopped the constant para...

Page 13: ...PowerPac Adaptor electrical connection voids EN61010 safety provisions E1 error code No load detected Verify all electrical connections displayed Verify buffer levels where appropriate Instrument not...

Page 14: ...ep 8 Caution Deactivating this safety feature increases the possibility that a failure of the power leads or loose output connection will not be detected E10 error code Unacceptable value s Clear the...

Page 15: ...r into its position Press the fuse holder gently until it snaps into place The unit is now ready for use Fig 5 Rear View Showing Fuse Drawer with Notches Note Repeated blowing of the fuse indicates a...

Page 16: ...5 min warmup at 300 V and 70 W Noise 55 dBA at 1 meter Readout Stability Volts 1 V Current 1 mA Safety Features No load detection Indicated by error message on display Sudden load change detection Ind...

Page 17: ...tors not specified by Bio Rad Laboratories for this power supply 4 Deliberate or accidental misuse 5 Damage caused by disaster For inquiry or request for repair service contact your local Bio Rad offi...

Page 18: ......

Page 19: ......

Page 20: ...rmany 089 31 884 0 Greece 30 210 9532 220 Hong Kong 852 2789 3300 Hungary 36 1 459 6100 India 91124 4029300 Israel 03 963 6050 Italy 39 02 216091 Japan 03 6361 7000 Korea 82 2 3473 4460 Mexico 52 555...

Reviews: