Maintenance and Troubleshooting

Error
Code

Explanation

Solution

E 6

Power Failure During a Timed Run

To clear the error code, press any key (other than

key.)

Power Failure detect (PFd) was 

The

key is disabled until the error code is cleared.

not activated; run was terminated 
either due to power failure or

Reminder

:

because the previous run was 

Time is reset to zero after each timed run.  Enter a time

stopped by turning off the power

value prior to each timed run.

supply or unplugging the unit.

If PFd is required, activate it prior to starting the run.

E 7

Power Failure During an Untimed Run

To clear the error code, press any key (other than

key.)

Run was terminated either due to 

The

key is disabled until the error code is cleared 

power failure, or because the 

from the display. 

previous run was stopped by 
turning off the power supply or

Reminder

:

unplugging the unit.

PFd can only be activated prior to a TIMED run.

E 9

Change in Load Resistance

Check and correct any potential resistance problems. Then,

• Electrophoresis cells were added 

• Press the

key to resume the run or,

or removed during a run

• To clear the error code, press any key (other than the

• Buffer leaking in a connected cell

key).

• Loose connection in a 

connected cell.

E 10

Unacceptable Value(s) Entered

• To clear the error code, press any key (other than

)

• No value entered

and enter acceptable values.

• PowerPac 200 voltage below 5 V

• When acceptable values are entered, press the

PowerPac 300 voltage below 10 V

key to resume the run.

• PowerPac 200 current below 0.01 A.

PowerPac 300 current below 4 mA.

E 11

Ground Leak Detected

• Check and correct the cell for an improper connection to 

(PowerPac 200 only)

earth ground.

• Restart the power supply by turning the power switch off 

and then on.

E3

Power Supply Regulation Error

Check for and correct problems, such as dirty contacts, 

E8

Possible power supply malfunction.

frayed wires, excessive buffer concentration, etc.  Then,

E12

• Press the

key, or 

E13

• Clear the error code by pressing any key (other than the 

E14

key).

E15

• If error code persists, 

note the error code number

and

E21

contact your Bio-Rad Representative.

E 98

Internal System Error

Contact your Bio-Rad Representative

E 99

Internal System Error

Contact your Bio-Rad Representative

Summary of Contents for PowerPac 200

Page 1: ...INSTRUCTION MANUAL AND Catalog Numbers PowerPac 200 165 5052 165 5053 PowerPac 300 165 5050 165 5051 POWER PAC 300 POWER PAC 200 CONST V mA V A POWER PAC 300...

Page 2: ...ion 7 Section 4 0 Maintenance and Troubleshooting 10 4 1 Maintenance 10 4 2 Troubleshooting 10 4 3 Replacing a Fuse 12 4 4 Firmware Version Number 12 Appendix A Specifications 13 Appendix B Warranty a...

Page 3: ...utlet using the 3 prong AC power cord provided with the power supply 3 Bio Rad electrophoresis cells have molded two prong plugs which are inserted into the power sup ply s high voltage output jacks T...

Page 4: ...cts are safe to use when operated in accordance with the instruction manual This safety certification does not extend to electrophoresis cells or accessories which are not I E C 1010 certified even wh...

Page 5: ...rements of 0 01 A Power output 200 watts maximum PowerPac 300 Voltage output Adjustable from 10 to 300 volts in increments of 1 volt Current output Adjustable from 4 to 400 milliAmps mA in increments...

Page 6: ...eatures including the following Constant voltage or constant current operation with automatic crossover Timer control Viewing angle adjustment 3 digit LED display Stackable case Automatic detection of...

Page 7: ...key down more than 1 sec changes values in 10V or 0 1 A steps PowerPac 300 Voltage and current values change in 1V or 1 mA steps respectively Holding either key down more than a few seconds changes v...

Page 8: ...e output terminals indicates high voltages may be present a Press the Power switch located on the side of the unit Press the side labeled l on the switch b The settings displayed are constant voltage...

Page 9: ...age limit is 200 V PowerPac 300 default voltage limit is 300 V a Press the Parameter key to select the Time parameter b Use the Scroll keys to enter the run time in minutes maxi mum 999 minutes Note D...

Page 10: ...Pac 300 PowerPac 200 Press the key The following options are available during the run To view the run conditions press the Parameter key To change the value of the constant parameter or the length of...

Page 11: ...e run or The current load dropped below To clear the error code press any key other than the the accepted rating as listed below key PowerPac 300 was less than 4 mA PowerPac 200 was less than 5 mA E 2...

Page 12: ...phoresis cells were added Press the key to resume the run or or removed during a run To clear the error code press any key other than the Buffer leaking in a connected cell key Loose connection in a c...

Page 13: ...e holder gently until it snaps into place The unit is now ready for use Caution Warning Disconnect the power supply before servicing Failure to follow this proce dure may result in personal injury and...

Page 14: ...to 2 00 A fully adjustable in 01 A steps PowerPac 300 5 mA to 400 mA fully adjustable in 1 mA steps Power max PowerPac 200 200 W PowerPac 300 75 W Terminals 4 pair recessed female banana jacks floatin...

Page 15: ...500 A Auto power up after power failure User selectable setup values maintained Safety Compliance I E C 1010 EMI Conforms to CE standards for Emissions and Immunity tested only at 220 V See Declarati...

Page 16: ...ad Laboratories B V Fokkerstraat 10 3905 KV Veenendaal Phone 0318 540666 Fax 0318 542216 New Zealand Bio Rad Laboratories Pty Ltd Unit 15 Poland Court 21 Poland Road P O Box 100 051 North Shore Mail C...

Page 17: ...______________________ ORDERING INFORMATION Catalog Number Product Description PowerPac 200 165 5052 PowerPac 200 100 120 V 165 5053 PowerPac 200 220 240 V PowerPac 300 165 5050 PowerPac 300 100 120 V...

Page 18: ...852 2789 1257 India Ph 91 124 6398112 113 114 Fx 91 124 6398115 Israel Ph 03 951 4124 Fx 03 951 4129 Italy Ph 34 91 590 5200 Fx 34 91 590 5211 Japan Ph 03 5811 6270 Fx 03 5811 6272 Korea Ph 82 2 3473...

Reviews: