BIO RAD BioLogic LP Quick Start Tutorial Download Page 2

1.

Buffer and Sample Preparation.

a.

Buffer “A”:  Empty the contents of the bottle labeled  Buffer “A” into a 500 ml graduated cylinder. Add degassed, deion-
ized water to 500 ml volume. Stir briefly. Label container as Buffer “A”.

b.

Buffer “B”: Prepare Buffer “B” and label container as described above.

c.

Sample: Dissolve the anion exchange standard in Buffer “A” using the syringe to measure 6.5 ml of buffer into the bottle
of protein standards. Replace the stopper and shake gently to dissolve. 

2.

Press the Program mode key; select New Method.

3.

Select Time programming mode. 

4.

Program the Pump.

a.

Press ADD.

b.

Select Buffer “A” with the Previous/Next keys; press OK.

c.

Enter the step length of 1 minutes; press OK.

d.

Enter flow rate of 3.0 ml/min; press OK.

e.

You have now entered the first step of the method (Buffer “A”, step length of 1 minute, flow rate of 3.0 ml/min.) Enter the
remaining steps.
Step 2. Gradient 0% to 50% “B”, step length 5 minutes, flow rate 3.0 ml/minute.
Step 3. Buffer “B”, step length 3 minutes, flow rate 3.0 ml/minute.
Step 4. Buffer “A”, step length 3 minutes, flow rate 3.0 ml/minute.

f.

After entering Step 4; press OK.

5.

Set the Alarm.

a.

Press the Alarm soft key.

b.

Press ADD.

c.

Enter a time of 1 minute for Alarm 1 and check that Hold Methods is set to NO.

d.

Press OK twice.
The alarm sounds after 1 minute to remind you to turn the valve back to the load position. If left in the Inject position, the
gradient will flow through the loop first instead of going directly to the column.

6.

Enter the fraction collector program.

a.

Press the Frac Coll soft key.

b.

Select ALL.

c.

Enter a fraction size of 0.5 minute.

d.

Press OK twice, then select DONE.

e.

Press SAVE, name the method “DEMO 1”, and press DONE.

7.

Load Sample loop with protein standard.

a.

Turn the MV-6 Injector valve knob counter-clockwise as far as it will go.

b.

Draw 2 ml of the protein standard mixture into the syringe.

c.

Insert the syringe in the top port of the MV-6 injector valve, and fill the sample loop. Leave the syringe in the port when
you have injected the sample — this prevents the sample from siphoning out of the loop.

8.

Start Method.

a.

Press the Run mode key. The system will count down 10 seconds, then start the method.

b.

When the method starts, turn the injector valve knob to the right as far as it will go. When the alarm sounds, turn the valve
knob to the left.

c.

If you are using a Bio-Rad 2128 fraction collector, press “Park”, then “Yes” when method is complete.

-0.50

0

4

6

8

10

12

1

0.00

0.50

1.00

1.50

2.00

2.50

3.00

3.50

4.00

4.50

5.00

0

10

20

30

40

50

60

70

80

90

100

Column: Econo Pac High Q Cartridge (Catalog #732-0028)

Buffer A: 25 mM Tris HCI, pH 8.1

Buffer B: 25 mM Tris HCI, pH 8.1 + 0.5 M NaCI

Standard: Anion Exchange Std. (Catalog #125-0561)

Equine myoglobin is not retained on the column — 

it elutes in the void volume (first peak). The remaining 

three proteins bind to the column and elute separately 

as the salt concentration increases.

A.U x1.E-2

mS/cm

TIME (MINUTES)

RUN CONDITIONS

2

3

4

5

6

7

8

9

10

11

12

E

s

Life Science
Group

Web site

www.bio-rad.com 

USA

(800) 4BIORAD 

Australia

02 9914 2800 

Austria

(01)-877 89 01 

Belgium

09-385 55 11

Brazil 

55 21 507 6191

Canada 

China

86-10-8201-1366/68

Denmark 

Finland

France

01 47 95 69 65

089 318 84-177

Hong Kong

852-2789-3300 

India

(91-124) 6398112/113/114

Israel 

03 951 4124 

Italy 

34 91 590 5200

Japan

03-5811-6270 

Korea

82-2-3473-4460 

Latin America

305-894-5950

Mexico

52 5 534 2552 to 54

The Netherlands

New Zealand

64-9-4152280 

47-23-38-41-30

Russia 

7 095 979 98 00

Singapore

65-2729877 

Spain

34-91-590-5200

Sweden

46 (0)8-55 51 27 00 

Switzerland

061-717-9555

United Kingdom

0800-181134

00-000    0000    Sig 0101

Bulletin 0000    US/EG    Rev A

Bio-Rad
Laboratories, Inc.

4106528 Rev A

Reviews: