3 INSTALLATION
Unpack the instrument from the box, taking care not to damage it as you unpack it.
1. Remove the strap and the adhesive tape
2. Remove the cardboard cover
3. Lift and remove the accessories and the internal cover
4. Remove the outer cardboard wall
5. To be able to lift the instrument, grip it only at the lower base of the housing, at the front or rear and pull it out
Installation
Place the instrument on a flat and stable work bench and connect it to the power outlet (230V ~ 50/60Hz) using the
cable supplied.
The stable, vibration-free laboratory table with horizontal top, vibration-free floor as far as possible.
The instrument must not be placed near the air outlet of an air conditioning appliance and must be protected from strong
light (window).
To ensure the full functionality of a heat sink, there must be a distance of at least 15 cm behind the instrument.
There must be no oils or chemical fumes near the work area.
Do not use extension cables or adapters and do not modify the cable supplied.