3) WATER AND ELECTRIC CONNECTIONS
Water inlet
The tap of the water sink (mixer) can supply hot and cold water while the tap of the central basin and the flexible
shower can supply only cold water.
The mixer is connected to N. 2 flexible pipes of
1/2” (one for hot water, the other one for cold water). They must be
connected to the laboratory’s water system using flexible pipes (not provided).
If the laboratory’s water system cannot supply hot water, it is necessary to close the provided hot water pipe using a
1/2” plug.
The hood is provided with valves of interception hot and cold water. In case of maintenance or failure it is possible to
close the valves in order to interrupt the dispensing of water to all the plumbing of the hood.
Water outlet
Near the water sink there is a
1”1/4 standard siphon. Except changes on request, it is placed on the back side of
the hood at ca. 40 cm from the ground (at ca. 20 cm with garbage disposal unit). Its position depends on the sink
placement.
Important:
The outlet system must have a minimum inclination of 1,5°.