16
Adapter #2022.
Humidifier --connected on "down-stream" side of SAFETY RELIEF
VALVE. Only a flow-through humidifier, having a low pressure drop, may
be used; otherwise protection provided by the SAFETY RELIEF VALVE is
rendered ineffective. A humidifier with a "bubbler" tube or pressure jet
should not be used.
Pressure Alarms (high and/or low) --T-connected to Gauge line.
A Humidifier or Pressure Alarm connection can affect system compliance.
Connecting hoses and water level in the Humidifier should be arranged so
that system compliance is increased minimally. This means that extra
hose lengths should be short, and water level in the Humidifier should be
maintained at a high level. Some alarm units affect the volume of gas in
the system. The operating characteristics of any auxiliary device should be
carefully considered before using it with the MVP-10.