- 28 -
Symptom: Red “-FS” or “+FS” light lit, won’t AutoZero
Problem/Cause
Solution/Corrective Action
Sensor wired
incorrectly
Check sensor wiring. Measure
resistance of each “arm” of bridge, if
possible; should be approximately the
same.
Bridge too far out of
balance as installed
Shunt one arm of bridge with resistor to
bring gage into balance (Contact
Binsfeld Engineering Technical Support.
Request BB900 Bridge Balancer.)
Symptom: Flickering “Data” light, spikes in data signal
Problem/Cause
Solution/Corrective Action
Unsteady battery
voltage at transmitter
Check transmitter power supply
(measure voltage at “+9V” relative to
“Gnd”). Check transmitter battery
connections.
Faulty RF
transmission path
Relocate receiving antenna to optimize
transmission path from transmitter.
Double check antenna cable
connections. Change RF channel (on
both transmitter and receiver).