MK / MKT (E5) 03/2019
page 40/158
7.
Set-point entry in “Fixed value” operating mode
In Fixed value operating mode you can enter a temperature set-point, the fan speed(with MKF 56), and
the switching-state of up to 16 operation lines.
All settings made in Fixed value operating mode remain valid until the next manual change. They are
saved also when turning off the chamber or in case of toggling to Idle Mode or Program Mode.
Setting ranges
Control ranges
Temperature
MK
-50 °C /
-58 °F
up to 180 °C /
356 °F.
(range -50 °C /
-58 °F
up to -40 ºC /
-40 °F
not provided for operation)
-40 ºC /
-40 °F
up to 180 °C /
356 °F
Temperature
MKT
-80 °C /
-112 °F
up to 180 °C /
356 °F
(range -80 °C /
-112 °F
up to -70 ºC / -
94 °F
not provided for operation)
-70 ºC /
-94 °F
up to 180 °C /
356 °F
Fan speed
MK 56
30 % up to 100 %
MK 56:
Reduce the fan speed only if required, because the spatial temperature distribution of
will also be reduced.
Technical data refers to 100% fan speed.
With set-point type “
Limit
”, adapt the safety controller (chap. 12.2) or the over/under tempera-
ture safety device class 2 (option, chap.12.3) always when you changed the temperature set-
point. Set the safety controller set-point or the set-point of the over/under temperature safety
device class 3.3 (option) by approx. 2 °C to 5 °C above the controller temperature set-point.
Recommended setting: Set-point type “
Offset
” with safety controller set-point 2 °C.
7.1
Set-point entry through the “Setpoints” menu
Press the
Setpoint setting
icon to access the “Setpoint” setting menu from Normal display.
“Setpoints” menu (example: MK 56).
Select “Fixed value operation setpoints” to access the
individual parameters.
•
Select the field “Temperature” and enter the desired temperature setpoint.
Setting range:
MK
-50 °C /
-58°F
up to 180 °C /
356 °F
,
MKT:
-80 °C /
-112 °F
up to 180 °C /
356 °F:
Confirm entry with
Confirm
icon.
•
MK 56:
Select the field “Fan” and enter the desired fan speed setpoint.
Setting range: 30% up to 100% fan speed. Confirm entry with
Confirm
icon.
Summary of Contents for MK 115
Page 138: ...MK MKT E5 03 2019 page 138 158 24 12 Dimensions MK 56 dimensions mm ...
Page 139: ...MK MKT E5 03 2019 page 139 158 MK MKT 115 dimensions mm ...
Page 140: ...MK MKT E5 03 2019 page 140 158 MK 240 dimensions mm ...
Page 141: ...MK MKT E5 03 2019 page 141 158 MKT 240 dimensions mm ...
Page 142: ...MK MKT E5 03 2019 page 142 158 MK MKT 720 dimensions mm ...
Page 144: ...MK MKT E5 03 2019 page 144 158 ...
Page 145: ...MK MKT E5 03 2019 page 145 158 ...
Page 146: ...MK MKT E5 03 2019 page 146 158 25 2 EU Declaration of Conformity for MKT ...
Page 147: ...MK MKT E5 03 2019 page 147 158 ...
Page 148: ...MK MKT E5 03 2019 page 148 158 ...
Page 150: ...MK MKT E5 03 2019 page 150 158 ...
Page 151: ...MK MKT E5 03 2019 page 151 158 26 Product registration 26 1 Registering a BINDER chamber ...
Page 159: ......