KBF-S (E6) 08/2018
page 27/88
4.3
Electrical connection
The chambers are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm /
70.87 in
in length.
Model version
Art. No.
(x = 0 or 1)
Power plug
V/-10% at the
indicated power fre-
quency
Current
type
Chamber
fuse
KBFS240-230V
KBFS720-230V
9x20-0366
9x20-0368
Grounded
plug
200-230 V at 50 Hz
200-230 V at 60 Hz
1N~
16 Amp
KBFS240UL-240V
KBFS720UL-240V
9x20-0367
9x20-0369
NEMA 6-20P
200-240 V at 50Hz
200-240 V at 60Hz
2~
16 Amp
•
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber’s protective conductor meets the lat-
est technology. The protective conductors of the socket and plug must be compatible!
•
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left chamber side, bottom right-hand, see chap. 1.4).
•
When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
•
Pollution degree (acc. to IEC 61010-1): 2
•
Installation category (acc. to IEC 61010-1): II
CAUTION
Danger of incorrect power supply voltage.
Damage to the equipment.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 24.4).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
Remark when operating the chamber with a power frequency of 60 Hz:
WARNING
High leakage current.
Electrical hazard.
Earth connection is essential before connecting supply. Check socket before inserting
plug.
When connected to a power supply 1N~ with a frequency of 60 Hz, a leakage current of more than 3.5
mAmp is possible. If grounding through the power cable is insufficient or missing, the leakage current
may flow through the user’s body. Correct installation of the professional grade power socket provided by
the user safely avoids this. Before connecting the chamber to the socket, please check its grounding
contact type plug for appropriate construction and if it is undamaged.
Summary of Contents for KBF-S 240
Page 75: ...KBF S E6 08 2018 page 75 88 24 7 Dimensions Dimensions size 240...
Page 76: ...KBF S E6 08 2018 page 76 88 Dimensions size 720...
Page 78: ...KBF S E6 08 2018 page 78 88...
Page 79: ...KBF S E6 08 2018 page 79 88...
Page 81: ...KBF S E6 08 2018 page 81 88...
Page 82: ...KBF S E6 08 2018 page 82 88 26 Product registration...
Page 89: ......