KBF-S (E6) 08/2018
page 71/88
24.3 Definition of usable volume
The usable volume illustrated below is calculated as follows:
c
c
b
a
C
B
A
b
a
A, B, C = internal dimensions (W, H, D)
a, b, c = distance to wall
a = 0.1*A
b = 0.1*B
c = 0.1*C
V
USE
= (A - 2 * a) * (B - 2 * b) * (C - 2 * c)
Figure 19: Determination of the useable volume
The technical data refers to the defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and humidity.
24.4 KBF-S technical data
Chamber size
240
720
Exterior dimensions
Width, net
mm /
inch
925 /
36.42
1250 /
49.21
Height, gross (incl. feet/castors)
mm /
inch
1460 /
57.48
1925 /
75.79
Depth, net
mm /
inch
800 /
31.50
890 /
35.04
Depth, gross (including door handle, I-
triangle, connection and 30 mm for
cable)
mm /
inch
880 /
34.65
970 /
38.19
Wall clearance rear (minimum)
(spacer)
mm /
inch
100 /
3.94
100 /
3.94
Wall clearance side (minimum)
mm /
inch
160 /
6.29
160 /
6.29
Doors
Quantity of doors
1
2
Interior dimensions
Width
mm /
inch
650 /
25.60
973 /
38.31
Height
mm /
inch
785 /
30.91
1250 /
49.21
Depth
mm /
inch
485 /
19.09
576 /
22.68
Interior volume
l /
cu.ft.
247 /
8.7
700/
24.7
Steam space volume
l /
cu.ft.
348 /
12.3
918 /
32.4
Summary of Contents for KBF-S 240
Page 75: ...KBF S E6 08 2018 page 75 88 24 7 Dimensions Dimensions size 240...
Page 76: ...KBF S E6 08 2018 page 76 88 Dimensions size 720...
Page 78: ...KBF S E6 08 2018 page 78 88...
Page 79: ...KBF S E6 08 2018 page 79 88...
Page 81: ...KBF S E6 08 2018 page 81 88...
Page 82: ...KBF S E6 08 2018 page 82 88 26 Product registration...
Page 89: ......