COMBIMASS
®
compact
Doc. No.: 0905ME20500
13
Chapter 2 Installation
Positioning the sensor rod, determining dimension B
Determining dimension B (without a feed unit)
Before installing the sensor, you should measure the length of the welded sleeve up to the inner wall of
the pipe (dimension "M" as on the left of the following diagram) in order to determine dimension "B".
Screw the threaded pipe nipple into the welded sleeve. Please use sealing material according to the spe-
cifications.
Screw the nut with the pressing ring and the cutting ring onto the threaded pipe nipple in such a way that
the sensor rod can still move.
Measure dimension "K" = reference edge up to sleeve (see on the left of above diagram).
Add dimensions "K" and "M" to obtain dimension "B".