3.2
Guidelines for safe lifting and transportation
After operation, please observe the guidelines for temporary decommissioning (chap. 23.2). Empty the
Permadry™ water pan before moving the chamber. In case of any spilling of the contents, shut down the
chamber and dry it out carefully and completely.
CAUTION
Sliding or tilting of the chamber.
Damage to the chamber.
Risk of injury by lifting heavy loads.
Transport the chamber in its original packaging only.
For moving or shipping, secure the chamber with transport straps.
∅
Do NOT lift or transport the chamber using the door handle, the door or the lower
housing.
Lift the chamber at the four lower corners with the aid of 4 people and place it on a
rolling pallet.
Move the chamber to the desired location and lift it from the rolling pallet with the aid of
four people.
•
Permissible ambient temperature range during transport: 10 °C /
14 °F
to +60 °C /
140 °F
.
You can order transport packing for moving or shipping purposes from BINDER service.
Permissible ambient temperature range during transport:
•
If the humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
.
With temperatures below +3 °C /
37.4 °F
, water must be completely removed from the humidifying
system.
CAUTION
Transport below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C /
37.4 °F
.
3.3
Storage
Intermediate storage of the chamber is possible in a closed and dry room. Observe the guidelines for
temporary decommissioning (chap. 23.2).
Permissible ambient temperature range during storage:
•
the humidifying system has NOT been emptied: +3 °C /
37.4 °F
to +60 °C /
140 °F
.
•
After BINDER Service has emptied the humidifying system: -10 °C /
14 °F
to +60 °C /
140 °F
With temperatures below +3 °C /
37.4 °F
, water must be completely removed from the humidifying
system.
CAUTION
Storage below +3 °C / 37.4 °F with filled steam humidifying system.
Freezing in the steam generator.
Damage to the chamber.
Contact BINDER Service before any transportation below +3 °C /
37.4 °F
.