C 150 (E2) 06/2011
page 47/78
10.7 CO
2
tolerance range alarm (CO
2
over/under concentration)
The main controller CO
2
alarm.
The CO
2
concentration has risen above or fallen below the CO
2
alarm threshold.
•
The alarm signal is usually emitted immediately upon the fault occurring, but no alarm signal is emitted
during the CO
2
alarm delay time after the outer door is closed.
•
Visual
display:
LED
Lower display, alternating
Percentage
CO
2
Alarm
code
Flashes
Flashes
•
Audible alarm: buzzer (intermittent sound)
•
Switching the zero-voltage relay alarm output
Actions:
•
Check the setting of the operating function “
AL.2
” (CO
2
alarm-threshold). The value should be
≥
1 vol.-
% CO
2
. If necessary, adjust the value.
•
Check that the outer door is closed properly.
•
Check the door's seals for any damage. Replace any damaged seals.
•
If the above points do not reveal the source of the fault, it may be that the unit is faulty. Please contact
BINDER Service.
In case of a
CO
2
over concentration
alarm, proceed as follows:
•
Open both unit doors for approx. 30 seconds.
•
Normal operation can be restarted, as soon as the requested values have been readjusted.
If the same alarm recurs, please contact BINDER Service.
10.8 CO
2
pressure too low
The CO
2
primary pressure at the intake valve is less than 0.3 bar /
4.4 psi
below the ambient air pressure.
The alarm displays show a pressure drop of CO
2
admission below 0.3 bar / 4.4 psi
•
Immediate alarm
•
Visual
display:
LED
Lower display, alternating
Percentage
CO
2
Alarm
code
Flashes
Flashes
•
Audible alarm: buzzer (intermittent sound)
•
Switching the zero-voltage relay alarm output