CB-S / CB-S-UL (E7) 06/2020
page 70/119
16.2.6 Assigning the DNS server address
Access to this function is possible if manual IP address assignment (chap. 16.2.1) or manual DNS server
address assignment (chap. 16.2.2) has been selected.
With manual IP address assignment:
Path:
Normal display
Settings
Ethernet
DNS server address
With manual DNS server address assignment:
Path:
Normal display
Settings
Ethernet
DNS server address
Press the
OK button
to enable the setting.
The DNS server address entry is done in four steps, corresponding to the number sections: (1).(2).(3).(4)
Principle of entry:
•
Use the
OK button
to select the desired section of the DNS server address 1/4, 2/4, 3/4, 4/4 in the
upper display line
•
Use the
Arrow buttons
to enter the value for the selected section of the DNS server address
•
For details please refer to the description of the similar procedure in chap. 16.2.3 “Assigning the IP
address”.
With the
Back button
you can go back to the “
Ethernet
” submenu and, repeatedly pressing it, to
Normal
display
.
17. Data recorder
An internal data recorder saves chamber data and events in three data sets.
With the export function “Export recorder data” (chap. 18.3) you can save the three data sets via the USB
interface to USB stick. They are issued in the selected language as a spreadsheet with the file extension
“.csv” and can be further processed in the desired program. The data is unencrypted. Always the entire
data memory is read out.
17.1 Recorded data
All data is given out in tabular form
. The headlines of the values “number”, “date”, and “time” are given out
in the selected language, all other information in English.
•
Chamber data for the user “DL1”
Tabular representation of the
actual temperature and CO
2
values together with the date and time, ac-
cording to the set storage rate (chap. 17.3). Temperature values are always given out in °C.
•
Chamber data for BINDER Service “DL2”
This data is intended for use by BINDER Service. It also contains information from the self-test func-
tion. The storage rate is fix (1 minute). Temperature values are always given out in °C.
•
Event list
Messages regarding the controller and data memory as well as the alarm messages
together with the
date and time:
•
Firmware update done
•
“New config (USB)”: New configuration uploaded via USB
•
“Data recorder cleared “ Data recorder and event list deleted via setup program
Summary of Contents for 9040-0189
Page 109: ...CB S CB S UL E7 06 2020 page 109 119 26 9 Dimensions CB S CB S UL 170 Dimensions in mm...
Page 110: ...CB S CB S UL E7 06 2020 page 110 119 CB S CB S UL 260 Dimensions in mm...
Page 112: ...CB S CB S UL E7 06 2020 page 112 119...
Page 113: ...CB S CB S UL E7 06 2020 page 113 119 28 Product registration...