12. Notifying and alarm functions
12.1 Notifying and alarm system overview (auto diagnosis system)
•
Visual indication of notifying or error message as a blue note on the screen of the controller MB1.
•
Visual indication of an alarm message as a red note with an alarm bell symbol.
Additionally there is an audible alert, provided that the buzzer has not been deactivated in menu “User
level” (chap. 6.4).
Event
Alarm
(red field)
Fault in refrigerating machine
FAULT COMPRESSOR
immediately
Operation line 1 (bedew protec-
tion) activated
DRY
immediately
Limit value of safety controller
exceeded
TEMPERATURE LIMIT
immediately
Power failure
With option over/under temperature safety device class 2 (chap. 11.3):
Exceeding the maximum / mini-
mum temperature
TEMP CLASS
immediately
With option keyboard locking (BINDER Individual, chap. 14.4):
Keyboard locked
KEY LOCK
immediately
The indicated intervals refer to the time after occurrence of the error or indicated condition.
12.2 Resetting the notifying or alarm messages
The RESET button which is used to acknowledge and reset the indication appears automatically when-
ever a note or an alarm message comes up.
1.
Depending on the type of error, eliminate the cause of the alarm or wait until the unit compensates for
the error.
2.
Press the RESET button to reset the note or the alarm message.
CAUTION
In case the RESET button does not cancel the notifying or alarm indication, the nature of
disturbance has not been removed correctly. The operation of the unit remains interrupted.
¾
Contact BINDER service.