MK (E3) 07/2009
page 24/70
6.4
Settings in the menu “User Level”
User Level
Date and time
Summer time
Language
Temperature unit
Buzzer
Safety controller
User-code No.
English
°C
1
Active
Date and time
Enter the actual date and time to provide the measurement records with the correct relation of time. Data
are displayed in the chart recorder function (chap. 7) of the controller and remain stored during power
failure.
Summer time
Time is set one hour in advance during the summer time period.
•
Summer time switch
•
Off:
No change to summer time occurs
•
User timed:
Beginning and end of summer time can be set individually
•
Automatic:
The summer time arrangement for central Europe is enabled
(summer time from last Sunday of March until last Sunday of October)
Language
Select the menu language as German, English, or French (chap. 6.1).
Temperature unit
Do NOT change the temperature unit from °C to °F.
Buzzer
•
Inactive:
No acoustic alarm in case of alarm events (chap. 12).
•
Active:
Acoustic alarm in case of alarm events (chap. 12).
Safety controller
Enter a safety controller tolerance limit for temperature exceeding. For setting, see chap. 11.3.
User-Code No.
Change the password (“user code”) needed to access the menu “User settings”. Factory default setting
+00001.
Keep in mind any change of the user password. There is no entrance to this menu without this
password.