3.2
Guidelines for safe lifting and transportation
The front castors can be blocked by brakes. Please move the units with castors only when empty and on
an even surface, otherwise the castors may be damaged. Mount the upholstered steel L-type profile at
the lower door locking. Observe the guidelines for temporary decommissioning (chap. 16.2).
CAUTION
Sliding or tilting of the unit.
Damage to the unit.
Risk of injury by lifting heavy loads.
¾
Transport the unit in its original packaging only.
¾
Secure the climatic chamber with transport straps for transport.
∅
Do NOT lift or transport the unit using the door handle, the door or the lower housing.
∅
Do NOT lift units by hand.
¾
Place units using technical devices (fork lifter) on the transport pallet. Set the fork lifter
only from the rear in the middle of the unit. Make sure to place all the lateral supports
of the unit on the forks.
∅
Transport units ONLY with the original transport pallet. Set the fork lifter only to the
pallet. Without the pallet the unit is in imminent danger of overturning.
•
Permissible ambient temperature range during transport: -10°C /
14°F
to +60°C /
140°F
.
You can order transport packing and pallets for transportation purposes from BINDER service.
3.3
Storage
Intermediate storage of the unit is possible in a closed and dry room. Observe the guidelines for tempo-
rary decommissioning (chap. 16.2).
•
Permissible ambient temperature range during storage: -10°C /
14°F
to +60°C /
140°F
.
•
Permissible ambient humidity: max. 70 % r.H., non-condensing
If following storage in a cold location the unit is transferred to the installation site for start-up, condensa-
tion may form. Wait at least one hour until the chamber has attained ambient temperature and is com-
pletely dry.
3.4
Location of installation and ambient conditions
Set up the temperature test chamber on an even and non-flammable surface, free from vibration, and in a
well-ventilated, dry location and align it using a spirit level. The site of installation must be capable of sup-
porting the unit’s weight (see technical data, chap. 18.4).
CAUTION
Danger of overheating.
Damage to the unit.
∅
Do NOT set up units in non-ventilated recesses.
¾
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation: +18°C /
64.4°F
to +32°C /
89,6°F
. At ele-
vated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +25°C /
77°F
to which the specified technical data relate. For other ambient condi-
tions, deviations from the indicated data are possible.