Calibrating and adjusting the temperature and humidity control
11.2 Calibrating and adjusting points
11.2.1 Calibrating and adjusting point of the temperature control
The adjustment of the temperature control is carried out as a single point adjustment.
Choose the most common operating temperature or a temperature situated in the middle of the most com-
mon used temperature range as calibrating and adjusting point.
11.2.2 Calibrating and adjusting points of the humidity control
The adjustment of the humidity control is carried out as a dual point adjustment.
Choose the values which limit the most common used humidity range as calibrating and adjusting points.
11.3 Calibrating and adjusting the temperature control
1.
Set the set temperature to the selected value (chap. 11.2.1).
2.
Set the set humidity to one of the selected values for the humidity calibration and adjustment (chap.
11.2.2).
3.
Reset the temperature controller adjustment.
4.
Calibrate the temperature controller (chap. 11.3.2).
5.
If necessary, adjust the temperature controller.
Service Manual KBF (E5.2) 04-2015