FDL (E2) 03/2011
page 51/68
13.2 Protocol printer (option)
The protocol printer is connected using the FDL interface via an interface converter RS 422 / RS 232.
Confirm that the interface mode is correctly set to “Printer” in the user level (chap. 10).
Figure 12: Protocol printer
Interface configuration:
Baud rate:
9600
Stop bit:
1
Parity:
none
The actual temperature values are put out regularly with fixed formatting.
Printout: one printed line for each print interval with relative time stamp, temperature value with one deci-
mal point, curve representation (see “Manual for Setting Matrix Printer Epson LX-300+“, Art. No. 7001-
0041). In every 5th line the set print interval “Ptime” in minutes is noted. The printer interval is set in the
user level (chap. 10).
Example:
Temp.: 34.7 *
Temp.: 35.6 *
Temp.: 32.8 *
Temp.: 30.1 *
Ptime : 001. *
Temp.: 27.4 *
Temp.: 26.9 *
Temp.: 26.6 *
Temp.: 26.4 *
Ptime : 001. *
Temp.: 26.2 *
In this example the print interval is set to 1 min, i.e., every minute a new temperature value is printed.
In conjunction with the specified print interval, real-time reference can be achieved by noting
down the starting time of recording.
For operating instructions of the printer, please refer to the operating manual of the printer
manufacturer.