40
DD1050
7
ERROR MESSAGES
7.1 Terminal errors
fault
Cause
Remedy
The terminal fails
to switch on when
the power-on key is
pressed
The blue light is off
Check the green light on the power
supplier:
if it is on, make sure that the connector is
plugged in correctly into point 9 (Fig. 3.1
The terminal does not
switch on
Wrong key pressed
Press on the logo in the center of the key
You will feel the key cap press down when
the key is depressed
Pressing on the touch
screen produces no
effect
Display not calibrated
Switch off the terminal and power it again.
Press on any part of the display within 5
seconds from powering.
Calibrate the display.
The double click on the
screen does not always
produce an effect
Wrong
double
click
timing
Switch off the terminal and power it again.
Press on any part of the display within 5
seconds from powering.
Calibrate the display.
7.2 Scale errors
fault
Cause
Remedy
9999999
The scale is overloaded
Reduce the weight to a value below the
maximum scale capacity so as to return
within the weighing zone.
- -
Weight cannot be reset Unload the scale.
- -
Weight instable
Eliminate the cause for weight instability
and attempt to power the terminal again.
- LOCK -
The instrument has not
been reset correctly
Unload the scale.