Caution
4. BEFORE USING THE DEVICE
4-1. Environmental settings
4-2. How to execute the zero adjustment
BEFORE USING THE DEVICE
If there is a difference in the temperature between the storage location and the utilization location,
leave the device for a certain amount of time in the utilization location to perform the average
environmental settings (temperature, humidity).
When the Level Master is turned on, depending on the utilization environment and on the status
of the reference surface measured, the level position may not be always the "0" position
(continuous blue LED). Execute every time the reference settings by means of zero adjustment
in accord with the utilization environment.
③
Push the POWER switch to turn ON the power
.
④
Execute the zero adjustment and the levelling checks following
the procedures
below.
②
Use as reference the line markers on the base of the device and
mark 4 points in the X and Y directions on the reference surface
.
4 markings
①
Position the device on the reference surface. When placing the
device, remove oil and dirt from its base and also remove notches,
oil and dirt from the reference surface of the precision plate.
(Example: If the difference in the temperature is 10°C, leave the device for 15-20 minutes.
)
How to check the levelling value in HIGH mode
Go to P6
→
How to execute the zeroadjustment in HIGH mode
Turn by 180°
②
The levelling value of both axes is within
0.08 after turning the device by 180°.
①
The levelling value of the X and Y axes
is within 0.08 after turning ON the power.
Go to P4
→
How to check the levelling value in LOW mode
②
The levelling value of one of the axes
exceeds 0.08 after turning the device by
180°The LED (red) is blinking.
①
The levelling value of the X and Y axes
is within 0.08 after turning ON the power.
Case1
Case2
Turn by 180°
Continuous LED (red)
Blinking LED (red)
How to check the levelling value in LOW mode
Go to P4
→
How to check the levelling value in LOW mode
①
The levelling value of one of the axes exceeds
0.08 after turning ON the power. The LED (red)
is blinking.
Case 3
Blinking LED (red)
Blinking LED (red)
Turn by 180
°
Case 2
Blinking LED (red)
③
The levelling value of both axes is
within 0.8 after turning the device by
180
°
.
②
The levelling value of the X and Y axes
is within 0.8 after switching to the
LOW
mode.
Turn by 180
°
Go to P5
→
How to execute the zero adjustment in LOW mode
Go to P5
→
How to execute the zero adjustment in LOW mode
Case 1
Continuous LED (red)
①
Switch to
LOW
mode.
③
The levelling value of one of the axes
exceeds 0.8 after turning the device by
180
°
.
②
The levelling value of the X and Y axes
is within 0.8 after switching to the
LOW
mode.
①
Switch to
LOW
mode.
Case 3
②
The levelling value of one of the axes
exceeds 0.8 after switching to the
LOW
mode.
The LED (red) is blinking.
①
Switch to
LOW
mode.
④
Adjust the level of the reference
surface in order to obtain a levelling
value within 0.8
for both axes.
Go to P5
→
How to execute the zero adjustment in LOW mode
③
Adjust the level of the reference
surface in order to obtain a levelling
value within 0.8
for both axes.
13
14
BEFORE USING THE DEVICE