6.4 Lubrication
6.4.1 Verifying cleanliness
Visually inspect the device to ensure it is
clean. If necessary, clean again using a
soft brush.
6.5 Sterilisation
CAUTION
The quality of the sterilisation is highly
dependent on how clean the in-
strument
is.
Only
perfectly
clean
instruments may be sterilised.
CAUTION
Do not use a sterilisation procedure
other than the one described below.
Procedure
Enclose the device and its accessories in
sterilisation bags which are large
enough to ensure the items can move
around, and which comply with the
standards in force (e.g.: EN 868-5). Ster-
ilise using steam on a class B cycle as
per EN 13060 / ISO 17665-1.
Note : All Bien- Air Dental SA in-
struments
are
sterilisable
in
an
autoclave up to 135°C. Duration: 3 or 18
minutes, depending on the national re-
quirements in force.
After cleaning, disinfecting and ster-
ilising the device, and before using it,
start it up at moderate speed with a
bur in the locking mechanism for 10 to
15 seconds to distribute and remove
the excess lubricant.
6.6 Servicing
Note : Bien-Air Dental SA recommends
that the user has his or her dynamic
devices checked or serviced every
three years.