Rx Only
CA 1: 2.5 L MS
ENG INSTRUCTIONS FOR USE
Other languages available onwww.bienair.com/ifu
REF
2100337-0002/2020.0
6
Page 1: ...Rx Only CA 1 2 5 L MS ENG INSTRUCTIONS FOR USE Other languages available on www bienair com ifu REF 2100337 0002 2020 06...
Page 2: ...REF 2100337 0002 2020 06 CA 1 2 5 L MS Bien Air Dental SA Set supplied REF CA 1 2 5 L MS 1601055 001 Optional accessories REF 1600617 006 1600036 006 1600064 006 1000001 010...
Page 3: ...r maintenance 10 6 1 2 Suitable maintenance products 10 6 2 Cleaning 11 6 3 Disinfection 13 6 3 1 Manual cleaning disinfection 13 6 3 2 Automatic disinfection 13 6 4 Lubrication 15 6 4 1 Verifying cle...
Page 4: ...ht or moderate injury or damage to the device if the safety instructions are not cor rectly followed Recyclable materials Use rubber gloves After initial mechanical resistance tighten fully in the dir...
Page 5: ...elating to medical devices Those medical devices comply with the legislation in force 2 3 Intended use Product intended for professional use only Used in general dentistry and oral surgery WARNING Any...
Page 6: ...n medical devices that are contaminated or potentially con taminated must comply with universal precautions in particular the wearing of personal protective equipment gloves goggles etc Pointed and sh...
Page 7: ...Technical data Contra angle 1 2 5 Standard coupling As per ISO standard 3964 Drive speed 40 000 rpm max Speed of rotation 100 000 rpm max Typical operating torque 1 5 Ncm Transmission ratio Red instru...
Page 8: ...m surgical burs the user must ensure by selecting the correct operating conditions that there is no danger to the user patient or third parties WARNING Follow the guidelines for use according to the b...
Page 9: ...ust not be started without a bur inserted Comply with maximum lengths by always inserting the bur as far as possible into the locking mech anism If a bur is operated at high speeds when incorrectly mo...
Page 10: ...ts Using other products or parts may cause faults during operation and or void the warranty CAUTION l Carry out cleaning disinfection sterilisation without a bur in the chuck mechanism l Do not immers...
Page 11: ...tap water at 15 C 38 C 2 Carefully remove all traces of dirt or deposits from the nozzles using the Bien Air cleaning wire optional 3 After selecting the appropriate end piece perform preliminary cle...
Page 12: ...al and external cleaning 1 After selecting the appropriate end piece spray the inside and the out side of the device with the product Spraynet FIG 6 and FIG 7 2 Dry with non woven compresses 12 FIG 6...
Page 13: ...rfaces with sterile non woven compresses im pregnated with sterile demineralized water 4 Eliminate the excess of disinfectant e g foam residue over the external surfaces of the device by spraying it w...
Page 14: ...0 C 95 C 194 F 203 F 5 10 minutes Drying 18 22 minutes CAUTION Never cool devices by rinsing them CAUTION If an automatic washer is used at the place of the washer thermo disinfector re spect the prev...
Page 15: ...on is highly dependent on how clean the in strument is Only perfectly clean instruments should be sterilised CAUTION Do not use a sterilisation procedure other than the one described below Procedure C...
Page 16: ...o its use CAUTION Comply with the expiration date of the sterilised items which depends on the storage conditions and type of pack aging 6 7 Servicing Note Bien Air Dental SA recommends that the user...
Page 17: ...s l Damage caused by unusual chem ical electrical or electrolytic influences l Faulty air water or electrical con nections CAUTION The warranty becomes null and void if damage and its consequences res...
Page 18: ...18 NOTES...
Page 19: ...ENG 19 NOTES...
Page 20: ...L nggasse 60 Case postale 2500 Bienne 6 Switzerland Tel 41 0 32 344 64 64 Fax 41 0 32 344 64 91 dental bienair com Other adresses available at www bienair com REF 2100337 0002 2020 06 CA 1 2 5 L MS B...