ENG
60
15.8 CA ENDO / Reciproc option
It is only possible to access the CA ENDO mode with the
iOptima
INT
system.
Tap
to enter the CA ENDO mode.
Predefined endo file systems foreseeing the reciprocating
movement need activation on the iOptima
INT
system.
In case the movement wasn’t previously activated, tapping it
leads to the activation page.
15.8.1 Activation page
A.
Tap
More information
and follow the instructions displayed.
B.
Input the code obtained and tap
Activate
(2).
A message box displays either:
C.
If activated the app navigates to the operative screen.
15.8.2 Operative screen
The pedal symbol (3) invites the user to press it.
A.
Press the dental unit pedal to switch ON the micromotor.
The movement symbol (4) displays the tool progression.
⚠
CAUTION
• Valid only with Bien-Air CA ENDO.
• CA ENDO must be used only with specific tools.
License not Activated
.
License Activated
.
FIG. 34
FIG. 35
FIG. 36
FIG. 37
FIG. 38
2
1
3
4
5
6
NOTES
24
New systems and brands can also be created directly from
the system screen.
25
Maximum storage of 10 user-defined brands and 10 systems
per brand, if reached the
icon disappears.
26
Restore ALL default system settings from the Brands screen.
27
It is only possible to delete user-defined brands
28
If an ENDO file systems foreseeing the reciprocating move-
ment is selected, speed, torque, motor direction, light, pedal mode
disappear and the gear ratio is fixed.
29
When the power demand of the MX2 micromotor is ex-
cessive, the operative screen displays the overheating symbol (5).
In this case the iOptima
INT
unit lowers the torque automatically in
order to avoid overheating of the MX2 micromotor. To restore
100% torque, allow the motor to idle or stop for a few seconds.
30
If stress is excessive on the file, the file stress alert dis-
plays
(6).
FILE
STRESS