background image

protherm II 

 

 

Observe  the  directions  for  use!  Handling  of  this  device  requires 
knowledge  and  adherence  to  these  instructions.  The  protherm  II  and 
accessories may only be used by qualified specialized staff. The patient 
must be monitored throughout the procedure. 

 

Fix the BIEGLER protherm II firmly to the infusion stand using the clamps at the 
back. Only use infusion stands or poles that are sufficiently stable. 

Connect  power  cable  to  power  supply.  Before  connecting  to  the  mains  power 
supply, check the voltage specified on the device label. The device gives a short 
beep and the standby light (Diag.2/7) lights up. 

If a different temperature to 38.5 °C is desired, it can be preset in Standby mode 
using the controls  

 and 

 (Diag.2/3 and Diag.2/4). If an adjustment control is 

pressed, the visual display indicates the existing preset temperature. By repeated 
operation  of  the  control 

  or 

  the  temperature  can  be  reset.  The  indicator 

automatically goes out after approximately 7 seconds. Resetting of temperature 
can only be performed in Standby mode. 

Heating of the protherm II can be started by pressing the control 

 (Diag.2/5). 

The protherm II attains the set target temperature within 1 minute. The indicator 
display shows the actual temperature (± 0.5 °C) as an illuminated band. 

Open  the  device:  release  the  two  clips  on  the  right  (Diag.1/6)  by  pressing  the 
button  on  the  clip  and  open  the  device.  The  heating  zones  are  visible,  the 
attachment hooks for the warming bag are on the reverse at the top (Diag.1/4). 

Hang  the  warming  bag  on  the  attachment  hooks  (Diag.1/4)  and  push  the  tube 
outlets into the tube holders (Diag.1/5, 7, 8). The tube with the two drip chambers 
must be placed at the top, the tube to the air trap below (Diag.1/3). The warming 
bag hangs without creases in the open device. 

Close the protherm II and lock with the clips on the right side (Diag.1/6). Place the 
air trap in the mounting provided (Diag.1/2). 

Fill the system. Take care that filling proceeds without air bubbles occurring.  

When the device is opened, the system must not be under pressure. Closing the 
system when it is under pressure can cause the warming bag to burst. The clips 
are designed so that they resist opening above a certain operating pressure. The 
device must not be forcibly opened. 

 

Summary of Contents for protherm II

Page 1: ...Instructions for use protherm II Blood and Infusion Warmer Edition 2017 11 21 220 240 V ENG...

Page 2: ...s and safety instructions as well as the operating instructions are read and carefully observed by all users before first using the device It is the responsibility of those using the device to fully a...

Page 3: ...cedure 7 3 2 Alarms 9 3 3 Shutting down the device 10 3 4 Troubleshooting 11 4 Maintainance 11 5 Cleaning and disinfection 12 6 Periodic inspections 12 7 Manufacturer liability 14 8 Warranty condition...

Page 4: ...be fastened to infusion stands or tripods which have sufficient stability and load capacity to support the device The device must only be used in areas in which the electrical installations are in acc...

Page 5: ...ed in conjunction with the protherm II Only use items with intact individual packaging Consumables are only for one time use disposable The reuse of disposable products results in possible infection r...

Page 6: ...y controlled heat zones The design of the casing allows rapid and simple fitting to any suitable infusion stand The plate temperature can be set between 37 C and 41 C in steps of 0 5 C and is displaye...

Page 7: ...for the air trap 3 Air trap 4 Attachment hooks for warming bag 5 Tube holder 6 Clips for closure of device 7 Tube holder 8 Tube holder Diag 2 1 Temperature scale 2 Temperature scale 3 Control to incr...

Page 8: ...d in Standby mode Heating of the protherm II can be started by pressing the control Diag 2 5 The protherm II attains the set target temperature within 1 minute The indicator display shows the actual t...

Page 9: ...The alarm can only be reset by pulling out the mains plug Wait for the temperature switch to cool down Visual signal Yellow alarm LEDs all green LEDs constantly on Yellow alarm LEDs all green LEDs con...

Page 10: ...alarm system does not have to be verified The alarm limits cannot be adjusted The risk management process has found that a shutdown of the alarms is not meaningful because it is desirable that the err...

Page 11: ...ion signal has been triggered The low temperature information signal is automatically cancelled after reaching 36 5 C Alarm LEDs light up audio alarm continuous tone Device was overheated Allow device...

Page 12: ...t such inspections based on their training knowledge and practical experience The results of the periodic inspection must be documented along with the date the inspecting agency and the device number...

Page 13: ...not triggered CHECKING THE HIGH TEMPERATURE ALARM Preheat the device to 41 C and wait for the temperature to stabilize then disconnect the mains plug Hold down the controls and and reconnect the main...

Page 14: ...within 24 months from the date of purchase will be repaired free of charge Claims are only accepted under the following terms The manufacturer and or supplier is informed immediately of the fault for...

Page 15: ...es assume liability for slight negligence The compensation for lost earnings and profits is likewise excluded 9 RETURN OF DEVICES Devices must be carefully cleaned and disinfected before being placed...

Page 16: ...guidance RF emissions acc to CISPR 11 Group 1 The protherm II uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nea...

Page 17: ...1000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interrupt...

Page 18: ...hs from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment m...

Page 19: ...to transmitter frequency m Rated maximum output power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 04 0 07 0 1 0 37 0 11 0 22 1 1 17 0 35 0 70 10 3 69 1 11 2 21...

Page 20: ...pliance with the Directive 93 42 EEC Observe the instructions for use Do not dispose of this product as unsorted municipal waste Control for increasing the temperature setting Control for decreasing t...

Page 21: ...tion of the batch Expiration date One time use only 14 OPERATING AND STORAGE CONDITIONS Permissible environmental conditions for the device and accessories Transport and storage Operating Temperature...

Page 22: ...Over temperature switch off 43 0 C 43 5 C 45 C 3 C Audio alarm frequency 3600 Hz sound pressure level 71 dbA Low temp information signal 500 ms 1 s sound pressure level 71 dBA Lowest liquid temperatu...

Reviews: