5
Only infusion bags that are capable of 300 mmHg must be used. Adhere to the
instructions for use for this bags.
The pressure infusion bags must be securely fastened at least 20 cm above the
patient to prevent air embolism. Prevent the connecting hose from kinking.
Always position the autopress in such a way that it is easy to operate without
obstacles. No other devices or infusion stands shall be positioned shortly before.
Make sure, that mains plug of the autopress is easily reachable to the operator.
(The mains plug is used to disconnect the device from mains)
The automatic pressure controller autopress is a class I ME equipment and
therefore only intended to be connected to supply mains with protective earth.
The autopress may not be used if:
the housing is damaged or one of the front film layers becomes detached
the device has been exposed to a hard physical shock (e.g. dropped, hit or
shaken)
the device has been immersed in water
the mains power cord or plug is damaged
the device has given someone an electric shock
the fixing clamp is damaged in such a way that safe clamping to the infusion
stand is no longer guaranteed.
The autopress may not be used if there is a malfunction:
e. g. display error, no pressure, …
If there is a malfunction, suitable warning signs should be attached to the device
to ensure that it is not used until the required service and repair work has been
carried out.
Safety Instructions for consumable materials:
Consumables are only for one-time use (disposable). The reuse of disposable
products results in possible infection risks for the patient or user.