SETTING THE TEMPERATURE UNITS 

1.  Press the Up arrow and Down arrow buttons together to toggle between °F and °C. 

 

Note:

 When the unit is tuned off, on subsequent power-up the units will revert to °C. 

SETTING THE OPERATING TEMPERATURE  

1.  To display the set temperature on the digital display, press and hold the set/

reset

 button. 

 

2.  To adjust the set temperature, press the set/

reset

 button and hold it while pressing the Up or 

Down  arrow  buttons.  Each  press  of  the  Up  arrow  or  Down  arrow  buttons  will  increase  or 

decrease the set temperature by 0.1°C. If the buttons are held down the temperature change 

will accelerate to 20°C per second. 

 

3.  When the set/

reset

 button is released, the current heater temperature is displayed. 

 

4.  The  heater  indicator  will  flash  if  the  new  set  temperature  is  higher  than  the  current  heater 

temperature and will continue to flash to show that the unit is controlling the temperature. 

 

5.  The  heater  indicator  will  turn  off  if  the  new  set  temperature  is  lower  than  the  current  heater 

temperature and will only start to flash when approaching the set temperature. 

 

6.  For optimal accuracy, allow the block inserts to equilibrate to the set temperature for about 30 

minutes. 

 

DB100/2TC 

The model  D100/2TC  accommodates  2  aluminium  blocks  with  independent  temperature  controls  so 

that each block may be set to a different temperature. The operation of the controls for each block is 

as described above. 

 

If using this model with only one block in place, the side with no block should have its set temperature 

adjusted to 30°C. 

 
Note:

  Due  to  variations  in  heat  losses  with  different  designs  of  insert  block,  the  actual  temperature 

may  vary.  The  units  are  calibrated  using  model  F3506  insert  blocks.  If  you  need  to  control  the 

temperature  to  a  greater  accuracy  than  the  instrument’s  set  point,  place  a  thermometer  in  a 

designated hole in one of the blocks. For greater accuracy still,  you can place a thermometer in the 

sample liquid in one of the test tubes.  

 

In either of these cases, it may be necessary to readjust the set temperature to achieve the precise 

temperature required. Allow the temperature of the unit to stabilise after each adjustment. There will 

be a time lag between the heater plate and the insert block achieving the set temperature due to the 

thermal contact between them.  

 

SETTING THE TIMER 

1.  To  set  the  timer,  press  and  hold  the  timer  button  and  press  the  Up  or  Down  arrow  buttons. 

Each press of the Up arrow or Down arrow buttons will increase or decrease the set time by 1 

minute. If the buttons are held down, the time change will accelerate to 200 min per second. 

 

2.  Release the timer button to activate the timer. An audible beep will be heard once the timer is 

activated. 

 

3.  When the count-down time has completed, an audible beep will be heard and the display will 

show  the  word  ‘End’  alternatively  with  the  current  heater  temperature.  Press  the  set/

reset

 

button to cancel the beeper. The unit  will start to cool to ambient once  the count-down time 

has completed. 

 

4.  To  stop  the  timer  at  any  point  during  the  count-down  period,  press  and  hold  the  set/

reset

 

button or the timer button until the word ‘Stop’ appears and an audible beep is heard. 

Summary of Contents for Dri-Block DB100/2

Page 1: ...DB100 2 DB100 3 DB100 4 DB100 2TC DB200 2 DB200 3 Dri Block Heaters OPERATOR S MANUAL Issue 7 06 2014 Part code 6106451...

Page 2: ...2...

Page 3: ...ALLAZIONE 13 Avvertenza 13 Sicurezza dell operatore 13 Installazione 13 Condizioni ambientali 14 Garanzia 14 INFORMACI N DE SEGURIDAD E INSTALACI N 15 Advertencia 15 Seguridad del operario 15 Instalac...

Page 4: ...4 ACCESSORIES 28 REPLACEMENT PARTS 29 DECLARATION OF CONFORMITY 30...

Page 5: ...an extraction tool for easy removal of the insert blocks The operating instructions in this Operator s Manual cover the following range of digital Dri Block heaters Model number Description Voltage fr...

Page 6: ...e unit please ensure that the following have been removed from the packaging Dri Block heater 6101308 Extraction tool Mains cables UK and EU or US Operator s manual Guarantee card The user is advised...

Page 7: ...re cut out If a safety problem should be encountered switch off the unit at the mains socket and remove the plug from the electricity supply INSTALLATION The instrument should be carried using both ha...

Page 8: ...s been tested for radio frequency interference and is certified under EN61326 GUARANTEE The instrument is guaranteed against any defect in material or workmanship for the period specified on the enclo...

Page 9: ...Bibby Scientific Ltd la protection pour l utilisateur ne serait alors plus garantie Tous les appareils Techne sont con us pour r pondre aux normes de s curit internationales et sont dot s d un coupe c...

Page 10: ...ort aux murs et autres quipements Remarque Les param tres sont indiqu s pour une temp rature ambiante de 20 C Ces caract ristiques peuvent se d t riorer en dehors d une temp rature ambiante de 10 30 C...

Page 11: ...m automatischen bertemperaturabschalter ausgestattet Wenn ein Sicherheitsproblem auftreten sollte mu das Ger t ausgeschaltet und vom Stromnetz getrennt werden INBETRIEBNAHME Das Ger t mit beiden H nde...

Page 12: ...lechtern Das Ger t wurde auf HF St reinfl sse gepr ft und entspricht den EMV Bedingungen nach EN61326 GARANTIE Techne gew hrleistet dass dieses Ger t f r den auf der Garantiekarte angegebenen Zeitraum...

Page 13: ...ilizzatore potrebbe essere a rischio Tutte le unit Techne sono state progettate in conformit ai requisiti internazionali di sicurezza e sono equipaggiate con un interruttore anti surriscaldamento Se s...

Page 14: ...i controllo sono indicate ad una temperatura ambiente di 20 C Le specifiche potrebbero peggiorare fuori da una temperatura ambiente compresa tra 10 C e 30 C Lo strumento stato collaudato per interfere...

Page 15: ...s requisitos internacionales de seguridad y traen incorporados un sistema de desconexi n en caso de sobre temperatura En caso de que surgiera un problema de seguridad desconecte el equipo de la red IN...

Page 16: ...0 C Las especificaciones pueden empeorar si se utiliza el instrumento fuera del intervalo de temperatura comprendido entre 10 C y 30 C Se han realizado pruebas para comprobar la interferencia de radio...

Page 17: ...at 37 C 1 0 C at 100 C 0 5 C at 37 C 1 0 C at 100 C 0 5 C at 37 C 1 0 C at 100 C Heat up time 25 C to 100 C 8 minutes 7 minutes 7 minutes Countdown Timer with audible beeps Yes Yes Yes Minimum Maximu...

Page 18: ...5 C at 100 C Heat up time 25 C to 100 C 8 minutes 7 minutes Heat up time 25 C to 200 C 23 minutes 19 minutes Countdown Timer with audible beeps Yes Yes Minimum Maximum timer 1min 99h59min 1min 99h59m...

Page 19: ...0 C at 100 C Heat up time 25 C to 100 C 8 minutes Overall dimensions W x D x H 279 x 260 x 105mm Countdown Timer with audible beeps Yes Minimum Maximum timer 1min 99h59min Safety over temp cut out The...

Page 20: ...containing the sample liquid in the blocks 4 The heater design temperature sensor and control circuit give good temperature control and uniformity but it is important that there is a close fit of the...

Page 21: ...indicator If this is lit it indicates that the unit has developed a fault and the over temperature cut out has been triggered This will cause the heater to be switched off and the unit will begin to...

Page 22: ...block should have its set temperature adjusted to 30 C Note Due to variations in heat losses with different designs of insert block the actual temperature may vary The units are calibrated using model...

Page 23: ...n We recommend that the blocks should be allowed to cool to 70 C or below before being removed from the heating unit To remove an insert block screw the extractor tool into the threaded hole and lift...

Page 24: ...rEAdy Measure the actual temperature of the block using the thermometer or probe 5 Press the set reset button and using either the Up arrow or the Down arrow buttons adjust the display to show the sam...

Page 25: ...the label next to the power inlet Note that fuses should only be replaced by a qualified electrician The holder for the two fuses is built into the mains input socket First remove the power cable and...

Page 26: ...Microfuge tubes 20 10 9 35 Tapered 95 x 75 x 50 F4464R 1 5ml Microfuge tubes red 20 10 9 35 Tapered 95 x 75 x 50 F4464B 1 5ml Microfuge tubes blue 20 10 9 35 Tapered 95 x 75 x 50 F4465 0 5ml Microfug...

Page 27: ...arts can be obtained from Bibby Scientific Ltd or your Techne dealer Product code Description HH179 S UK mains lead with plug 230V HH180 S European mains lead with plug 230V FCABLEUS US mains lead wit...

Page 28: ...30 DECLARATION OF CONFORMITY...

Page 29: ...l 41 41 541 5877 Fax 41 41 541 5878 Email General Information Sales info kesn ch Accounts Department accounts kesn ch USA O ce 233 Rogue River Highway 425 Grants Pass Oregon 97527 5429 USA Tel 1 503 2...

Reviews: