Cleaning of the optic outlet

Clean the optic outlet using isopropyl alcohol 

or other disinfection agents.

Blow the optic outlet dry with air syringe or dry 

it carefully with a soft cloth.

Avoid scratching the optic outlet! 

Never use a cleaning agent that has an 

ammonium, chlorine, or acid base. Carry 

out a visual inspection after each cleaning 

process.

1.

2.

1

2

Summary of Contents for Maxso X99 Series

Page 1: ...Maxso X99 Series 1 5 Electric High Speed Handpiece Instruction For Use...

Page 2: ...age General explanations without risk to persons or objects Thermal washer disinfectable Sterilizable up to the stated temperature Symbols 2 Introduction 3 4 Before Use 5 Product description 6 Technic...

Page 3: ...your patients Prior to initial use please read the instructions for use These explain how to use your Beyes product and guarantee a smooth and efficient operation Intended use The dental contra angle...

Page 4: ...nce with these Instructions for Use The handpiece should be prepared for use and maintained by staff who have been trained in procedures for infection control personal safety and patient safety Improp...

Page 5: ...sue risk of burning due to the press button heating up Use only non contaminated filtered oil free and dry compressed cooling air to operate the Maxso handpiece The operation of the contra angle is pe...

Page 6: ...6 Product description Contra angles 1 Press button 2 Spray nozzle 3 Compact optic outlet 4 Grip profile 5 retractable lug only contra angles with light 1 5 3 2 4 3...

Page 7: ...st be set at the same time Contra angles X99 Series Transmission ratio 1 5 Motor coupling acc to standard ISO 3964 Instrument acc to EN ISO 1797 1 1995 mm 1 6 Length in mm approved by Beyes mm 25 Min...

Page 8: ...ing the operation Push the contra angle handpiece onto the motor and turn it until it engages audibly Check the connection to the motor is secure Remove the contra angle handpiece from the motor by pu...

Page 9: ...Insert rotary instruments only when the contra angle handpiece is stationary Do not interfere with the running or slowing down of rotary instruments Do not active the rotary chuck of the handpiece or...

Page 10: ...time insert the rotary instrument until it reaches the limit stop 1 b Rotary instrument shank 2 35 mm Press the button 1 a and at the same time insert the rotary instrument and turn it until it snaps...

Page 11: ...s e g vibrations unusual noise overheating coolant supply failure or leakage stop the handpiece immediately and contact an authorized service partner organization Always check that the bur is secure b...

Page 12: ...thing Remove the rotary instrument Remove the contra angle handpiece from the motor Clean and disinfect the contra angle handpiece immediately after every treatment to flush out any liquid e g blood s...

Page 13: ...o proteinfixing effects Manual cleaning internal and external Rinse under demineralized water 38 C with the aid of a brush brush is not recommended in U K Then remove any liquid residues absorbent clo...

Page 14: ...remove dirt and deposits Clean and disinfect the nozzle cleaner in an ultrasonic bath disinfection bath Clean coolant tubes Blow through the coolant tube with an air syringe In the case of clogged spr...

Page 15: ...or other disinfection agents Blow the optic outlet dry with air syringe or dry it carefully with a soft cloth Avoid scratching the optic outlet Never use a cleaning agent that has an ammonium chlorine...

Page 16: ...ertified by officially recognized institutes For USA Use EPA registered surface disinfectants Note the manufacturer s specifications for the use of the disinfectants After manual cleaning disinfection...

Page 17: ...processing with thermo washer disinfector Note the manufacturer s specifications on equipment cleaning agents and detergents Make sure that the contra angle is completely dry internally and externall...

Page 18: ...ctions for use of the oil spray can or Only with Beyes Caremaster See instructions for use of Caremaster Recommended lubrication cycles Essential after every internal cleaning Before each sterilizatio...

Page 19: ...ervice of the chuck system Only with Beyes Lube Fit the spray nozzle onto the spray can Hold the contra angle handpiece firmly Press the tip of the spray nozzle firmly into the chuck system Spray for...

Page 20: ...ownwards Start the contra angle handpiece for 30 seconds to remove excess oil Start at the minimum speed and increase to them maximum speed within 5 to 10 seconds Repeat the entire hygienic maintenanc...

Page 21: ...Follow the operating instructions of the manufacturer Clean disinfect and lubricate prior to sterilization Wrap the contra angle handpiece and the accessory in sterile goods packing according to EN 86...

Page 22: ...rilizers in accordance with EN 13060 Sterilization holding time a minimum of 3 minutes at 134 C 273 2 F or Steam sterilization class S with sterilizers in accordance with EN 13060 The sterilizer manuf...

Page 23: ...er no circumstances be liable for incidental or consequential damages or damage due to improper use or maintenance This Warranty is subject to the following conditions Beyes will without charge repair...

Page 24: ...law of the states in which he or she practices to use or order the use of this device Lotus NL B V Koningin Julianaplein 10 Ie Verd 2595AA The Hague Netherlands Tel 31645171879 English 31626669008 Du...

Reviews: