TROUbLEShOOTING

1: The system is still working, but the air volume is weak and the suction force is greatly 

reduced

Reason: 

The air inlet is blocked or there is foreign matter in the tube, and the filter reach

-

es saturation and lose efficacy then forms a blockage alarm Check: Turn off the power, 

open the top cover, check if there is any foreign matter blocking the air inlet; then check 

if the air outlet is blocked. If the suction of the empty machine is normal, the filter element 

needs to be replaced.

2: The machine automatically shut down immediately after starting the machine, turn it on 

again and repeat the above issue

Reason: 

The internal airflow is not smooth, the filter resistance is too big, causing the 

fan starting current to be too big, the machine will shutdown to automatically protect the 

machine

Check:

 Turn off the power, open the upper cover, check if there is any foreign matter 

blocking the air inlet; if it is checked, after the treatment is normal, adjust the air volume to 

the minimum, then turn it on again, and adjust the air volume to the appropriate position.

3: Turn on the power switch, the indicator light is off

Check

 if the socket is in poor contact. If the power socket is intact, check if the fuse in the 

power box is burnt. If it is damaged, please replace the fuse

Be sure to cut off the power before checking and repairing the equipment! 

Summary of Contents for DuraMax X1

Page 1: ...X1 Instruction for Use Leading aerosol control system to protect dental professionals from COVID19...

Page 2: ...2 CONTENT Safety precautions 3 PACKING LIST 3 Features 4 SPECIFICATIONS 4 Parts number 5 operation 6 7 INSTALLATION INSTRUCTIONS 7 FILTERS replacement 8 Troubleshooting 9 WARRANTY 10 11...

Page 3: ...e wall outlet before cleaning the unit and when not using the unit Turn off all controls before unplugging Do not insert fingers or foreign objects into the intake or air outlet The batteries must be...

Page 4: ...th a handy remote control for smooth operation Specifications Model DuraMax X1 Voltage AC110V 60Hz Power 350W Suction 8 2 Kpa Speed options 10 Ultraviolet C UV C 254nm Spectrum 10W Maximum volume 7500...

Page 5: ...5 Parts number...

Page 6: ...Press 3 to decrease time 5 Press 4 to confirm XXX represents the number of hours the machine is running during timing the maximum is 300 hours the minimum is 1 hour After setting press the time butto...

Page 7: ...increase time 4 Press 3 to decrease time 5 Press 4 to confirm INSTALLATION INSTRUCTIONS 1 Place the main body of the unit at a convenient spot connect the arm onto the main body 2 Connect the power c...

Page 8: ...Primary Filter Eliminate droplet aerosol and dust Replace every month Stage 2 Secondary Filter Eliminate blood and aerosol Replace 3 month Stage 3 HEPA Filter Eliminate virus bacteria size 0 3 m 99 97...

Page 9: ...turn it on again and repeat the above issue Reason The internal airflow is not smooth the filter resistance is too big causing the fan starting current to be too big the machine will shutdown to autom...

Page 10: ...30 months 3 EXCLUSIONS This limited warranty does not cover and BEYES shall not be liable for the following 1 defects damage or other conditions caused in whole or in part by misuse abuse negligence a...

Page 11: ...ective parts Beyes shall not be liable for and hereby disclaims any direct special indirect incidental exemplary or consequential damages or delays including but not limited to damages for loss of pro...

Page 12: ...el 1 855 603 1888 Fax 1 855 720 1228 Email info beyes ca Website www beyes ca Printed in Canada File ENI026 Rev1 30 04 20 Lotus NL B V Koningin Julianaplein 10 Ie Verd 2595AA The Hague Netherlands Tel...

Reviews: