background image

11

Mains connection 230 V~; 50 Hz; 230 VA

UK: 240 V~; 50 Hz; 240 VA

Saudi Arabia: 220 V~; 60 Hz; 220 VA

Expected 

service life

1000 h

Operating 

conditions

Temperature: +10 °C to +40 °C

Relative humidity: 10% to 95%

Ambient pressure: 700 to 1060 hPa

Storage and 

transportation 

conditions

Temperature: 0 °C to +60 °C

Relative humidity: 10% to 95%

Ambient pressure: 500 to 1060 hPa

Aerosol properties 1) Flow: 5.3 l/min

2) Aerosol delivery: 0.326 ml

3) Aerosol delivery rate: 0.132 ml/min

4) Particle size (MMAD): 3.07 µm

The serial number is located on the device or in the 

battery compartment.

Subject to technical changes.

Particle size diagram

Measurements were performed using a sodium fluoride 

solution with a “Next Generation Impactor” (NGI).

The diagram may therefore not be applicable to sus-

pensions or highly viscous medications. You can obtain 

more detailed information from the manufacturer of 

your medication.

12.  Replacement parts and 

wearing parts

Designation

Material

REF 

IH 26 Kids year pack (contains 

mouthpiece, silicone children’s 

mask, silicone baby mask, 

atomiser, compressed air 

hose, filter)

PP/PVC/

silicone

601.19

Standard IH 26 year pack 

(contains mouthpiece, nose-

piece, adult mask, children's 

mask, atomiser, compressed 

air hose, filter)

PP/PVC

601.28

Nasal douche

PP

601.37

Baby mask

PVC

601.31

Removable “giraffe”

PP/ PVC

164.182

 Note

If the device is not used according to the instruc-

tions specified, perfect functionality cannot be gu-

aranteed! We reserve the right to make technical 

changes to improve and develop the product. This 

device and its accessories comply with the Eu-

ropean standards EN60601-1 and EN60601-1-2 

(CISPR 11, IEC61000-3-2, IEC61000-3-3, 

IEC61000-4-2, IEC61000-4-3, IEC61000-4-4, 

IEC61000-4-5, IEC61000-4-6, IEC61000-4-7, 

IEC61000-4-8, IEC61000-4-11) as well as EN13544-1 

and are subject to particular precautions with regard 

to electromagnetic compatibility. This device meets 

the requirements of European Directive 93/42/EEC for 

medical devices, as well as those of the Medizinpro-

duktegesetz (German Medical Devices Act).

NOTES ON ELECTROMAGNETIC COMPATIBILITY

• The device is suitable for use in all environments li-

sted in these instructions for use, including domestic 

environments.

• The use of the device may be limited in the presence 

of electromagnetic disturbances. This could result in 

issues such as error messages or the failure of the 

display/device.

• Avoid using this device directly next to other devices 

or stacked on top of other devices, as this could lead 

to faulty operation. If, however, it is necessary to use 

the device in the manner stated, this device as well 

as the other devices must be monitored to ensure 

they are working properly.

13. Warranty and service

Beurer GmbH, Söflinger Straße 218, 89077 Ulm, Ger-

many (hereinafter referred to as “Beurer”) provides a 

warranty for this product, subject to the requirements 

below and to the extent described as follows.

 

The warranty conditions below shall not affect 

the seller’s statutory warranty obligations which 

ensue from the sales agreement with the buyer. 

The warranty shall apply without prejudice to any 

mandatory statutory provisions on liability.

Beurer guarantees the perfect functionality and com-

pleteness of this product.

The worldwide warranty period is 5 years, commencing 

from the purchase of the new, unused product from 

the seller.

Summary of Contents for IH 26 Kids

Page 1: ...IH 26 Kids EN Nebuliser Instructions for use...

Page 2: ...Ensure you are calm and relaxed when inhaling the medication and breathe slowly and deeply to ensure that the medication reaches right down to the fine lower bronchi Breathe out normally Once it has b...

Page 3: ...sponsible for their safety or who are instructed by such a person in how to use the device Check the leaflet of the medication for any contrain dications for use with the usual systems for aerosol tre...

Page 4: ...e for the purpose for which it is designed aerosol inhalation and in the manner specified in these instructions for use Any form of improper use can be dangerous In the event of an acute emergency the...

Page 5: ...me Note Clean and disinfect the atomiser and accessories before using them for the first time See Cleaning and disinfection on page 8 The adapter 17 with the Removable giraffe 16 can optionally be att...

Page 6: ...dication container 11 Ensure that the connection is correct 6 4 Inserting the valve Press the valve 10 into the opening intended for this purpose on the upper side of the atomiser 8 Note The valve sys...

Page 7: ...n 21 Saline container 22 Cover hole Starting the application You can use the nasal douche to clean your nasal ca vities The nasal douche creates a fine jet of liquid for treating respiratory tract pro...

Page 8: ...the cap has been removed take the filter out of the device e g with tweezers or similar 2 Re insert the filter cap 5 with a new filter 3 Ensure that it is securely in place 8 Cleaning and disinfectio...

Page 9: ...nfecting the individual parts on a daily basis after the last usage as a minimum measure All you need for this is a little white vinegar and dis tilled water First clean the atomiser and accessories a...

Page 10: ...nce the atomiser starts to make a notably different sound What should be taken into account when using the device with infants and children 1 On infants and children the mask should cover the mouth an...

Page 11: ...h the Eu ropean standards EN60601 1 and EN60601 1 2 CISPR 11 IEC61000 3 2 IEC61000 3 3 IEC61000 4 2 IEC61000 4 3 IEC61000 4 4 IEC61000 4 5 IEC61000 4 6 IEC61000 4 7 IEC61000 4 8 IEC61000 4 11 as well...

Page 12: ...following are explicitly excluded from this warranty deterioration due to normal use or consumption of the product accessories supplied with this product which are worn out or used up through proper u...

Reviews: