54
10. Error messages
Error
message
Problem
Solution
Measurement during self-test, device not yet
ready for measurement.
Wait until the forehead symbol stops
flashing.
Room temperature below 15 °C or above
40 °C (<59 °F, >104 °F).
Room temperature must be between
15 °C and 40 °C (59 °F - 104 °F).
(1) Forehead temperature mode: The deter-
mined temperature is higher than 42.2 °C
(108 °F).
(2) Object temperature mode: The determined
temperature is higher than 80 °C (176 °F).
Operate the thermometer only with-
in the specified temperature ranges.
In the event of a repeated error mes-
sage, contact your retailer or Custom-
er Services.
(1) Forehead temperature mode: The deter-
mined temperature is lower than 34 °C
(93.2 °F).
(2) Object temperature mode: The determined
temperature is lower than 0 °C (32 °F).
Operate the thermometer only with-
in the specified temperature ranges.
In the event of a repeated error mes-
sage, contact your retailer or Custom-
er Services.
The batteries are empty.
Replace the batteries.