Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
The [KD-551] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [KD-551] should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic
material,
the
relative humidity should be
at least 30 %.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
typical
commercial
or
hospital
environment.
24
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The [KD-551] is intended for use in the electromagnetic environment specified below. The
customer or the user of the [KD-551] should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601test
level
Compliance
level
Electromagnetic environment -
guidance
Radiated RF
IEC 61000-4-3
3 V/m 80 MHz
to 2.5 GHz
3 V/m
Portable
and
mobile
RF
communications equipment should be
used no closer to any part of the
[KD-551], including cables, than the
recommended separation distance
calculated
from
the
equation
applicable to the frequency of the
transmitter.
Recommended separation distance
:
80 MHz to 800 MHz
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
P
d
2
.
1
P
d
3
.
2
manufacturer
and
d
is
the
25