Bestcare BESTLIFT PL400EF Manual Download Page 6

5

WARNING!

•  NO

 modification of this equipment is allowed. It is unsafe to modify the 

 

equipment from its original state. This product is deemed safe to use with  

only the original parts as delivered. Unauthorized use of any modified 

 

equipment may result in serious injury or death.

•  DO NOT 

replace any components of the lift without consulting with  

Bestcare, and must follow proper instruction from Bestcare when replacing  

any components.

•  DO NOT

 service any parts of the lift while in use with a patient.

•  It is 

NOT

 safe to load the equipment more than the indicated maximum safe  

working load.

•  IF

 any accessories with lower load capacity than the equipment itself is used, 

DO NOT

 load the equipment more than the lowest maximum safe  

working load.

•  REMOVE

 the battery pack if the lift will not be used for some time, to prevent  

battery leakage. 

PERIODICALLY 

check the battery pack and the charger

•  This equipment 

DOES NOT

 consist of any materials or ingredients to  

which patients are exposed, that may cause bodily harm or injury. 

•  Prolonged exposure or physical contact with any part of the equipment 

WILL  

NOT

 cause skin irritation, allergic reaction, etc.

•  Using accessories, detachable parts, or materials not described in the  

instruction manual 

MAY RESULT IN SERIOUS INJURIES.

•  Keep the equipment away from children and pets when in use. 

IT MAY  

CAUSE SERIOUS INJURY OR DEATH.

•  Keep the equipment away from any pests or pest infested area. Diseases  

transferred from pests to the equipment 

MAY CAUSE SERIOUS ILLNESS

  

to patients upon contact.

•  If the lift is exposed to the sun for long periods of time, be 

CAUTIOUS

  as the  

surface of the lift may be hot.

•  There are no risk if the lift is exposed to reasonably foreseeable  

environmental conditions

•  CAUTION!

 The controller contains parts and assemblies that could be  

susceptible to 

damage by electrostatic discharge (ESD)

. To reduce  

the likelihood of this, one should be grounded prior to using the controller.

•  DO NOT

 use the lift on an inclined surface that is greater than 5°.

•  CAUTION! 

The black plastic bolt caps can fall off the bolt, and they are a 

choking hazard

.

 

If the cap will not stay on the bolt, replace it immediately.

Summary of Contents for BESTLIFT PL400EF

Page 1: ...o l d a b l e P o r t a b l e Ensure the product has been assembled according to the instructions in this manual All operators should read and understood the instructions for safe and proper operation...

Page 2: ...manufacturing date can be found on the serial number label that is located at the top of the mast The YY in the serial number is the manufactured year and the MM is the manufactured month MAX LOAD 40...

Page 3: ...never have to be worried about your lift not being charged and ready for use The PL400EF is intended to be used with patients who are completely immobile or totally dependent on the patient lift for...

Page 4: ...physical contact with the patient in order for the device to carry out its intended Do Not Iron Item may not be smooted or finished with an iron Class II symbol To identify equipment is designed that...

Page 5: ...wise severely handicapped The patient lift should be used solely for transferring a user patient from one utility beds chairs toilets etc to another The patient lift should not be used for transportin...

Page 6: ...xposed that may cause bodily harm or injury Prolonged exposure or physical contact with any part of the equipment WILL NOT cause skin irritation allergic reaction etc Using accessories detachable part...

Page 7: ...t 8 Standing Bracket 13 Emergency Lowering 4 Leg Opening Pedal 9 Emergency Stop Button Charger 0 6 lb 0 3 kg 5 Base 10 Control Unit 11 lb 5 kg Combined Weight 1 5 69 8lb 31 7 kg UNLESS OTHERWISE SPECI...

Page 8: ...I L M mm in A 1841 2 72 5 B 1237 5 48 7 C 603 7 23 8 D 526 7 20 7 E 812 8 32 0 F 771 4 30 4 G 576 6 22 7 H 336 4 13 2 I 544 1 21 4 J 258 4 10 2 K 491 0 19 3 L 121 0 4 8 M 20 3 0 8 N 1198 8 47 2 O 657...

Page 9: ...mm I Base Opened Turning Diameter 55 1384 9mm H Base Closed Turning Diameter 51 1306 5mm Unloaded Lift Weight 89 7 lb 40 8 kg Standard Features 6 Point Spreader Bar 4 Rear Caster with Brake 3 Front Ca...

Page 10: ...g the mast Step 4 Lift the mast upright Lock the rear casters first then lift the mast upright by grabbing onto the handle bars Step 2 Secure the stand to the mast Make sure the tab on the standing br...

Page 11: ...10 Step 5 Insert the pin to the bottom hole to secure the mast in upright position Step 6 Remove the pin that is securing the boom Step 7 Remove lower actuator pin...

Page 12: ...ll the lower end of the actuator as follows It is critical to have the motor part of the actuator to face the same direction as the big arrow as follows Step 9 Install the upper end of actuator and re...

Page 13: ...ND CONFIDENTIAL THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF BESTCARE LLC ANY REPRODUCTION IN PART OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF BESTCARE LLC IS PROHIBITED 9 27...

Page 14: ...rage or Transport Step 1 Spreader bar at lowest position Step 2 Folding the boom Step 3 Folding the mast Step 4 Standing position 1 1 2 2 3 3 Warning The lift may fall and cause bodily injury if it is...

Page 15: ...one that serves the needs of the patient while providing the patient with optimal safety Never interfere with the lift unless instructed by the attendant Have a doctor nurse or medical attendant expe...

Page 16: ...mproperly Keep long charger cord out of walkway where it poses a tripping hazard Have the charger port on the controller facing away from the wall for easy disconnection When removing the battery need...

Page 17: ...ring charging the LCD Display Panel will show one of the four signs below The DC Charger must be connected The emergency stop button must be released The battery will not be charged if the emergency b...

Page 18: ...17 LED Indicator Green In Use Blank Standby Lifting Lowering Operating Lift Using Hand Control Operating Lift Using LCD Display Panel...

Page 19: ...w the LCD Display Panel will show the sign on the left either when the emergency stop button is released or when a button on the LCD Display Panel is pressed Lift operation is deactivated Battery need...

Page 20: ...ment with Head Support Polyester M L XL Liko Guldmann Replacement Polyester with PU Coating M L XL Liko Guldmann Replacement with Head Support Polyester with PU Coating M L XL Liko Guldmann Disposable...

Page 21: ...1 Hook up F Point 2 Hook up E Insert F through E 8 Strap Sling A B G H C D E F Point 1 Hook up A C Point 2 Hook up B D Point 3 Hook up E Point 4 Hook up F Point 5 Hook up H Point 6 Hook up G Insert H...

Page 22: ...d Transfer From Wheelchair to Bed 1 Grasp the sling at each corner of the U shape of the commode aperture 2 The sling should be fitted with the grab handles on the back section facing outward 3 Help t...

Page 23: ...0 years or 20 000 cycles The expected lifetime of the electrical components are 3 years Contact Bestcare for any recycling information First Received Monthly Every 3 Months Check connections between b...

Page 24: ...propyl alcohol For slings with rigid supports remove the plastic inserts before washing Look for an opening near the plastic inserts for removal Machine wash warm or cold Do not wash with other colors...

Page 25: ...g on usage The actuator will either go up or down but not both Replace your hand control as it is at the end of its life cycle Your control box may be malfunctioning There is a grinding sound inside t...

Page 26: ...he user The Emergency Lowering Device is located at the top of the actuator shaft It is intended for use if the actuator or electronics fail to operate while the patient is suspended in midair The dev...

Page 27: ...e date for the specified time period WARRANTY PERIOD PRODUCT OR COMPONENT Three 3 Years Patient lift frame or spreader bar Two 2 Years Actuator control box pendant charger weigh scale Stand Aids all c...

Page 28: ...Bestcare LLC 2920 Pacific Drive Norcross GA 30071 Ver 2021 09 www bestcarellc com 2021 Bestcare LLC...