2.4 Properties Description
Property
Text
Description
Name
English letters or
Numbers
Name of the logger
SN
10 Characters
Must be the same as the one in the label of
logger.
Sampling Interval (s)
Number from 1 to 240
Sampling frequency in LCD screen
Logging Interval (s)
Number from 2 to 86400
Logging frequency.
Logs
The count of records in the memory.
Capacity
The total capacity of the logger’s storage.
Start Mode
Start Now
Start log when the Logger was turn ON.
Delay Start
Wait Delay Time then start log.
Timing Start
Start log when Delay Time: HHMMSS
e.g. 170000 means 17:00:00
Delay Time
Input a Delay Time or Timing Time.
High & Low Limit
CH1: Temperature
CH2: Humidity
When the value over the limit, the buzz will
sounds an alarm.
Offset
4
Input positive number to decrease the value.
Input negative number to increase the value.
4
Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications.
However, there may be times when you wish to adjust the calibration of your logger. BSTSoftware
provides you with the ability to perform a single point offset calibration. This calibration can be used
to increase the accuracy of the logger for a restricted data range.