20
WWW.BERTIN-INSTRUMENTS.FR
USER MANUAL
Starting/stopping collection
Once the device has been started up and the single-use assembly installed
correctly, use the arrows to navigate the main menu and select collection mode,
then confirm with the “PLAY” button.
The “RESET” button returns you to
the previous menu. Pressing and holding the
down arrow locks or unlocks the keypad, to prevent unintentional operations
during the collection (unintentionally stopping the collection process, etc.).
Figure 9: Collection mode
1
Immediate collection:
Starts collection without programming, stops at user
request.
2
Simple collection
: Starts a collection delayed by an interval T1 and stops
after a programmed period T2.
3
Advanced collection
: Starts a collection delayed by an interval T1, stops
after a programmed period T2 and then restarts after a timed delay T3,
performed over N cycles (T1+T2).
Note: For the immediate sampling, the screen continuously displays the time
elapsed from the start of the collection throughout the entire collection process
For the simple and advanced program it displays the time remaining since the start
of the collection process.