4 Installation
LB 8901 Shield
18
64449BA2 Rev. 01, 10/2018
4.6
Installation of the shield
To avoid unnecessary exposure to radiation, install the shield as final system
component.
Tip
If the shield contains a radioactive source, the measurement of the
background count rate of the detector should take place before the shield
is installed.
j
CAUTION
Danger caused by ionising radiation
After removal of the transport cap, the beam path is free and ionizing
radiation occurs.
Consult the radiation safety officer responsible for your company.
Install the unit with at least two persons (mo holding / screwing).
Remove the transport cap only shortly before installation.
Hold the flange of the shield only in the direction of the mounting
location.
1
2
3
Padlock
Screws with lock washers
Transport cap
Fig. 5
Installation of the shield
1.
Place the shield in the mounting position.