5. Contamination Measurements
43
If the nuclide or nuclide mix to be measured is not yet included
in the small nuclide table, proceed as follows
Select
Nuclides
on the
System Menu
.
Figure 4.11: System menu
Press the
Enter
key. The nuclide table is displayed. The nu-
clides contained in the small nuclide table are marked with
an X.
Figure 4.12: Nuclide table. Pre-selected nuclides are
marked by an X.
Move the cursor to the nuclide you wish to include in the
small nuclide table using the
↓↑
key, and then press the
key to go to the next function level. Now press the
Edit
but-
ton. A new menu will appear listing all nuclide parameters:
Name, Active, Radiation Type, Mass Number, Unit, Cal.
Factor A-100
(according to DIN 25415)
, Cal. Factor ISO
(according to DIN ISO 7503-1)
, Alarm Threshold, Integral
Threshold, Integral Unit
and
Time Base.
Now move the
cursor to
Active
to select the nuclide with
Enter
.
Press
ESC
twice to quit the nuclide table. You get into the
System Menu
.