LB 470 Level
7 Main Menu Device Setup
56925BA2 Rev. 03, 11/2017
113
Adjust: Low Level Adjust
Device settings | Setup | Calibration | Adjust
After calibration, a level adjustment can be performed. A level adjustment must be
performed if the level shown is not the actual level. The lower adjustment can only
be performed at a level < 50%. The count rate at 100% will be kept fixed, while all
other points of the curve will be adjusted according to the adjusted count rate at
the entered level value.
1
2
Adjusted curve (linear calibration type)
Adjusted curve (exponential calibration type)
Fig. 93
Adjusted curves (Low Level Adjust)
NOTICE
Make sure that the source is mounted and the beam path is open. The container
must be empty, or be below the limit value.
1
2
3
4
Button
< Perform adjustment >
Input field
″
Process value
″
Input field
″
Measuring Duration
″
Process bar
5
6
7
8
9
Button
< Ok >
Display field
″
Count rate
″
[cps]
Button
< Cancel >
Button
< Start >
Button
< Stop >
Fig. 94
Lower adjustment
1.
Click on the field Process value (Fig. 94, item 3) to open the input field.
2.
Specify a percentage value for the lower adjustment and confirm with the
Enter key.
3.
Click on the
″
Read-In-Time
″
field (Fig. 94, item 2) and specify the duration of
measurement in seconds. The higher you set the measurement time, the
Summary of Contents for Duo Xpert
Page 2: ......
Page 4: ......
Page 119: ...LB 470 Level 7 Main Menu Device Setup 56925BA2 Rev 03 11 2017 119...
Page 181: ...LB 470 Level 12 Appendix 56925BA2 Rev 03 11 2017 181...
Page 184: ......
Page 190: ...1 A propos de ce manuel d utilisation LB 470 Niveau Level 8 56925BA59 Rev 01 11 2017...
Page 196: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 00 11 2017...
Page 227: ...Technical Information Level Technische Information F llstand 56925TI1L Rev 00 11 2017...